Start Up and Operating Control
Page 4-8
NOTE
Image Intensifier Field Size
Field Size
Press the
“
Field Size
”
key to select the X-Ray field size. The illuminated LED indicates which field size is
selected. Press the
“
Field Size
”
key until the field size you want is selected. Refer to the table below:
TEXT ICON FIELD SIZE
NORM
9 inches
MAG1
6 inches
MAG2
4.5 inches
Collimator Control
The X-Ray beam may be collimated by using the iris collimator or the shutter collimator. Press the key until
the shutter or iris is in the position you desire.
Shutter Collimator Rotation
Press the left key to rotate the shutter collimator counter-clockwise. Press the right key to rotate the
shutter collimator clockwise.
Shutter Collimator Open/Close
Press the
“
Shutter Collimator Open/Close
”
keys to open and close the shutter collimator. Press the bottom
portion of the key to close the shutters and press the top portion of the key to open the shutters.
Iris Collimator Open/Close
Press the
“
Iris Collimation Open/Close
”
keys to open or close the collimator iris. Press the bottom portion of
the key to close the iris or the top portion of the key to open the iris.
When the system is transferred to film mode, Iris opens fully. And the parameters set in the fluoroscopy
mode are memorized. When return to fluoroscopy mode, it can be restored until a reset is made.
Proper use of the iris can decrease the X-Ray radiation to the patient and improve image
quality.
Содержание Brivo OEC 715
Страница 9: ...Chapter1 Introduction and Safety...
Страница 34: ...Chapter2 System Overview...
Страница 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Страница 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Страница 40: ...Chapter3 Mechanical Positioning...
Страница 52: ...Chapter4 Start Up and Operating Control...
Страница 77: ...Chapter5 Usual Operations...
Страница 108: ...Chapter6 SpecialApplications...
Страница 139: ...Chapter7 Radiographic Film...
Страница 142: ...Chapter8 System Setup...
Страница 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Страница 150: ...Chapter9 Maintenance...
Страница 160: ...Chapter10 Display Messages...
Страница 162: ...Chapter11 Labels and Symbols...
Страница 167: ...Labels and Symbols Page 11 6 18 1 System rating plate indicates manufacture information and input power requirements...
Страница 168: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 11 7 19 1 X ray Generator rating plate Total Filtration label...
Страница 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Страница 179: ...Chapter12 Technical Reference...
Страница 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Страница 194: ...Technical Reference Page 12 16 Lateral configuration 1 5m LOHMANN Lateral configuration 1m LOHMANN...
Страница 206: ...Technical Reference Page 12 28 12 11 System Block...
Страница 209: ......
Страница 210: ...www gehealthcare com Imagination at work...