![GE Brivo OEC 715 Скачать руководство пользователя страница 153](http://html1.mh-extra.com/html/ge/brivo-oec-715/brivo-oec-715_operators-manual_4193902153.webp)
Maintenance
Page 9-4
NOTE
9.2.3.
Emergency Switch Performance Check
This check is performed as a matter of routine operator maintenance to ensure that the Emergency Switch
safety feature is functioning properly. Perform this check as often as possible in conditions such as removal
from long-term storage and exposure to fluids.
1.
Press and hold the vertical column control key and press Emergency Switch while motion is occurring.
2.
Verify that mechanical motion stops and the X-Ray Off indicator lights up.
3.
Press X-Ray switch and verify that X-Rays are disabled.
4.
Press the power on/off switch to power off the system.
5.
Rotate the Emergency Switch upward and then reboot the system.
Images will be lost when the system is restarted unless you have saved them. If you have
saved images you can access them through the W
orkstation’s Image Directory functi
on.
9.2.4.
Fluoro Mode Performance Check
This procedure produces X-Rays. Take the appropriate precautions.
1.
Position the C-Arm so that the tube is directly above the image intensifier.
2.
Place a suitable test object on the image intensifier.
3.
Verify that auto mode is selected. The LED on the top left corner of the
“
Auto Technique
”
key on the
control panel will be illuminated.
4.
Press the left X-Ray switch on footswitch and handswitch. Verify that a digitized fluoro image of the
object appears on the left monitor and that it remains there after the exposure is terminated.
5.
Verify operation of the field size selection keys: NORM, MAG1 and MAG2.
6.
Press the footswitch or handswitch to verify operation of the motorized collimation controls: shutter
rotation, iris collimation, and shutter open/close.
WARNING
Содержание Brivo OEC 715
Страница 9: ...Chapter1 Introduction and Safety...
Страница 34: ...Chapter2 System Overview...
Страница 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Страница 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Страница 40: ...Chapter3 Mechanical Positioning...
Страница 52: ...Chapter4 Start Up and Operating Control...
Страница 77: ...Chapter5 Usual Operations...
Страница 108: ...Chapter6 SpecialApplications...
Страница 139: ...Chapter7 Radiographic Film...
Страница 142: ...Chapter8 System Setup...
Страница 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Страница 150: ...Chapter9 Maintenance...
Страница 160: ...Chapter10 Display Messages...
Страница 162: ...Chapter11 Labels and Symbols...
Страница 167: ...Labels and Symbols Page 11 6 18 1 System rating plate indicates manufacture information and input power requirements...
Страница 168: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 11 7 19 1 X ray Generator rating plate Total Filtration label...
Страница 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Страница 179: ...Chapter12 Technical Reference...
Страница 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Страница 194: ...Technical Reference Page 12 16 Lateral configuration 1 5m LOHMANN Lateral configuration 1m LOHMANN...
Страница 206: ...Technical Reference Page 12 28 12 11 System Block...
Страница 209: ......
Страница 210: ...www gehealthcare com Imagination at work...