Standby or Power Off
Standby:
1.
Turn the C-Arm key switch to the standby position (
). The system is still power on.
Placing the equipment in standby disables X-Rays and vertical column operation so these
features will not be activated unintentionally.
2.
Turn the key switch back to on position when you are ready to use the C-Arm.
Power off:
1.
Press power on/off key on Workstation to power off system.
2.
Unplug the Workstation power cord from the AC receptacle.
3.
Disconnect the Workstation interconnect cable from the C-Arm.
When you turn the system off, wait for at least 1 minute before powering the system back
on. This helps prevent from operational problems.
Remove all power from the system before moving the system or if a problem occurs which
prevents normal operation. All power should also be removed when periodic maintenance
and cleaning is performed.
During shutting down, the power shall not be cut off automatically until the system
software exits.
During powering off, the system will ignore power on command by pressing power on/off
button until shutdown finished.
NOTE
NOTE
Содержание Brivo OEC 715
Страница 9: ...Chapter1 Introduction and Safety...
Страница 34: ...Chapter2 System Overview...
Страница 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Страница 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Страница 40: ...Chapter3 Mechanical Positioning...
Страница 52: ...Chapter4 Start Up and Operating Control...
Страница 77: ...Chapter5 Usual Operations...
Страница 108: ...Chapter6 SpecialApplications...
Страница 139: ...Chapter7 Radiographic Film...
Страница 142: ...Chapter8 System Setup...
Страница 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Страница 150: ...Chapter9 Maintenance...
Страница 160: ...Chapter10 Display Messages...
Страница 162: ...Chapter11 Labels and Symbols...
Страница 167: ...Labels and Symbols Page 11 6 18 1 System rating plate indicates manufacture information and input power requirements...
Страница 168: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 11 7 19 1 X ray Generator rating plate Total Filtration label...
Страница 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Страница 179: ...Chapter12 Technical Reference...
Страница 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Страница 194: ...Technical Reference Page 12 16 Lateral configuration 1 5m LOHMANN Lateral configuration 1m LOHMANN...
Страница 206: ...Technical Reference Page 12 28 12 11 System Block...
Страница 209: ......
Страница 210: ...www gehealthcare com Imagination at work...