![GE Brivo OEC 715 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/ge/brivo-oec-715/brivo-oec-715_operators-manual_4193902017.webp)
Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 1-9
1.6.3.
Dose to Patient
Skin Dose Levels
The skin dose levels in normal use may be high enough to cause deterministic effects. The availability of the different settings has a considerable
effect on the radiation quality, the delivered dose rate and the image quality.
Patient Entrance Reference Point
Air Kerma (Kinetic Energy Released per unit Mass, i.e. Dose) is an indication of the radiation delivered to the patient entrance reference point, a
point in space, and is measured in the units of Grays (Gy). The System determines the patient entrance reference point as a point along the central
X-ray beam axis 70cm from the focal spot (30cm from the image intensifier entrance).
Selection of the patient entrance reference point is based on international convention and typical usage of a multi-use mobile
fluoroscopy system. This point should represent a typical intersection of patient surface and the X-ray beam axis.
Radiation Data
The measurement is performed at patient entrance reference point. Following dosimetric data are typical values in normal use cases. There might
be with some tolerance in actual use cases.
The following data are for Brivo OEC 715, 785 with I.I.(23XZ4ST/BS).
Operation
Mode
Technique
Auto (PMMA 200mm)
Max
Pediatrics
With Grid, No Filter
With Filter, No Grid
With Grid, No Filter
With Filter, No Grid
Mag mode
Mag0
Mag1
Mag2
Mag0
Mag1
Mag2
Mag0/1/2
Mag0/1/2
Normal
Fluoro
kVp
80
81
80
72
73
74
110
110
mA
2.1
3
4
1.7
2.5
3.6
4
4
dose rate
(mGy/min)
10.33
15.17
19.67
3.36
5.15
7.70
38.87
24.29
High Level
Fluoro
kVp
81
81
80
72
73
74
110
110
mA
3.2
4.6
6.4
2.5
3.7
5.4
6
6
dose rate
(mGy/min)
16.18
23.26
31.48
4.94
7.62
11.56
58.30
36.43
NOTE
Содержание Brivo OEC 715
Страница 9: ...Chapter1 Introduction and Safety...
Страница 34: ...Chapter2 System Overview...
Страница 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Страница 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Страница 40: ...Chapter3 Mechanical Positioning...
Страница 52: ...Chapter4 Start Up and Operating Control...
Страница 77: ...Chapter5 Usual Operations...
Страница 108: ...Chapter6 SpecialApplications...
Страница 139: ...Chapter7 Radiographic Film...
Страница 142: ...Chapter8 System Setup...
Страница 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Страница 150: ...Chapter9 Maintenance...
Страница 160: ...Chapter10 Display Messages...
Страница 162: ...Chapter11 Labels and Symbols...
Страница 167: ...Labels and Symbols Page 11 6 18 1 System rating plate indicates manufacture information and input power requirements...
Страница 168: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 11 7 19 1 X ray Generator rating plate Total Filtration label...
Страница 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Страница 179: ...Chapter12 Technical Reference...
Страница 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Страница 194: ...Technical Reference Page 12 16 Lateral configuration 1 5m LOHMANN Lateral configuration 1m LOHMANN...
Страница 206: ...Technical Reference Page 12 28 12 11 System Block...
Страница 209: ......
Страница 210: ...www gehealthcare com Imagination at work...