Introduction and Safety
Page 1-22
Guidance and manufacturer’s declaration
- Electromagnetic Immunity
The
Brivo OEC 715/785/865 Mobile C-arm X-ray Product is intended to use in the electromagnetic environment
specified below. The purchaser or operator of the Brivo OEC 715/785/865 Mobile C-arm X-ray Product should assure
that it is used in an electromagnetic environment as described below:
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power
(W)
Distance
(m)
I
MMUNITY
TEST LEVEL
(V/m)
385
380
–
390
TETRA 400
Pulse
modulation
b)
18 Hz
1,8
0,3
27
450
430
–
470
GMRS 460,
FRS 460
FM
c)
5 kHz deviation
1 kHz sine
2
0,3
28
710
704
–
787
LTE Band 13,
17
Pulse
modulation
b)
217 Hz
0,2
0,3
9
745
780
810
800
–
960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0,3
28
870
930
1 720
1 700
–
1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
b)
217 Hz
2
0,3
28
1 845
1 970
2 450
2 400
–
2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
2
0,3
28
5 240
5 100
–
5 800
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0,2
0,3
9
5 500
5 785
NOTE If necessary to achieve the
IMMUNITY TEST LEVEL
, the distance between the transmitting antenna and the
ME EQUIPMENT
or
ME SYSTEM
may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a)
For some services, only the uplink frequencies are included.
b)
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent
actual modulation, it would be worst case.
Содержание Brivo OEC 715
Страница 9: ...Chapter1 Introduction and Safety...
Страница 34: ...Chapter2 System Overview...
Страница 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Страница 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Страница 40: ...Chapter3 Mechanical Positioning...
Страница 52: ...Chapter4 Start Up and Operating Control...
Страница 77: ...Chapter5 Usual Operations...
Страница 108: ...Chapter6 SpecialApplications...
Страница 139: ...Chapter7 Radiographic Film...
Страница 142: ...Chapter8 System Setup...
Страница 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Страница 150: ...Chapter9 Maintenance...
Страница 160: ...Chapter10 Display Messages...
Страница 162: ...Chapter11 Labels and Symbols...
Страница 167: ...Labels and Symbols Page 11 6 18 1 System rating plate indicates manufacture information and input power requirements...
Страница 168: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 11 7 19 1 X ray Generator rating plate Total Filtration label...
Страница 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Страница 179: ...Chapter12 Technical Reference...
Страница 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Страница 194: ...Technical Reference Page 12 16 Lateral configuration 1 5m LOHMANN Lateral configuration 1m LOHMANN...
Страница 206: ...Technical Reference Page 12 28 12 11 System Block...
Страница 209: ......
Страница 210: ...www gehealthcare com Imagination at work...