Start Up and Operating Control
Page 4-24
Is
“
Auto Brightness/Contrast
”
key
turned off?
Turn on
“
Auto
Brightness/Contrast
”
key.
When the automatic function is
selected in Normal Fluoro mode,
does technique reach maximum
limitation? (110kVp)
Switch to HLF mode or Digital
Spot mode.
Does air cover large portion in the
image?
Re-position to decrease the
portion of air in the image, or use
collimator leaf or Iris to cover the
portion of air.
Is the smart metal level too high?
Change the smart metal level
manually. When there is no metal
in the field, adjust the smart
metal level to
“
0
”
.
3
Image looks just
grayish (Too small
contrast)
Is
“
Auto Brightness/Contrast
”
key
turned off?
Turn on
“
Auto
Brightness/Contrast”
key or
manually adjust the contrast
level.
4
Image looks too
black-and-white (Too
high contrast)
Is
“Auto Brightness/Contrast”
key
turned off?
Turn on
“
Auto
Brightness/Contrast”
key or
manually adjust the contrast
level.
Does air cover large portion in the
image?
Re-position to decrease the
portion of air in the image, or use
collimator leaf or Iris to cover the
portion of air.
Is the smart metal level too high?
Change the smart metal level
manually. When there is no metal
in the field, adjust the smart
metal level to “0”.
5
Image looks too noisy
Is the image acquired in Pulsed
mode?
Switch to continuous mode or
prolong exposure time.
Is the image acquired in Low Dose
mode?
Switch to continuous mode.
Is the
“
Noise Filter
”
set too low?
Increase the noise filter setting.
Is the
“
Edge Enhancement
”
level
set too high?
Decrease edge enhancement
level setting.
6
Image looks blurry
(details are not clear)
Patient or C-Arm was moving
during the acquisition?
Decrease the noise filter setting.
Содержание Brivo OEC 715
Страница 9: ...Chapter1 Introduction and Safety...
Страница 34: ...Chapter2 System Overview...
Страница 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Страница 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Страница 40: ...Chapter3 Mechanical Positioning...
Страница 52: ...Chapter4 Start Up and Operating Control...
Страница 77: ...Chapter5 Usual Operations...
Страница 108: ...Chapter6 SpecialApplications...
Страница 139: ...Chapter7 Radiographic Film...
Страница 142: ...Chapter8 System Setup...
Страница 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Страница 150: ...Chapter9 Maintenance...
Страница 160: ...Chapter10 Display Messages...
Страница 162: ...Chapter11 Labels and Symbols...
Страница 167: ...Labels and Symbols Page 11 6 18 1 System rating plate indicates manufacture information and input power requirements...
Страница 168: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 11 7 19 1 X ray Generator rating plate Total Filtration label...
Страница 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Страница 179: ...Chapter12 Technical Reference...
Страница 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Страница 194: ...Technical Reference Page 12 16 Lateral configuration 1 5m LOHMANN Lateral configuration 1m LOHMANN...
Страница 206: ...Technical Reference Page 12 28 12 11 System Block...
Страница 209: ......
Страница 210: ...www gehealthcare com Imagination at work...