Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 1-19
Guidance a
nd manufacturer’s declaration
- Electromagnetic Immunity
The
Brivo OEC 715/785/865
Mobile C-arm X-ray Product is intended to use in the electromagnetic
environment specified below. The purchaser or operator of the
Brivo OEC 715/785/865
Mobile C-arm
X-ray Product should assure that it is used in an electromagnetic environment as described below:
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
(Ed3.0)
6 kV contact
8 kV air
(Ed3.0)
Floors are wood, concrete, or
ceramic tile, or floors are
covered with synthetic material
and the relative humidity is at
least 30 %.
8 kV contact
15 kV air
(Ed4.0)
8 kV contact
15 kV air
(Ed4.0)
Electrical fast
transient Pulse
IEC 61000-4-4
2 kV for power
supply lines
1 kV for
input/output lines
2 kV for power supply
lines
1 kV for input/output
lines
Mains power quality is that of a
typical commercial and/or
hospital environment
Surge
IEC 61000-4-5
1 kV Line(s) to
Line(s)
2 kV Line(s) to earth
1kV Line(s) to Line(s)
2 kV Line(s) to earth
Mains power quality is that of a
typical
commercial
and/or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
(250/300 cycle)
0% UT
0% UT
Mains power quality should be
that of a typical commercial or
hospital environment. If the
operator of the Brivo OEC
715/785/865 Mobile C-arm
X-ray Product requires
continued operation during
power mains interruptions, it is
recommended that the Brivo
OEC 715/785/865 Mobile C-arm
X-ray Product be powered from
an uninterruptible power
supply or a battery.
Power frequency
magnetic field
(50/60Hz)
IEC 61000-4-8
(50Hz or 60Hz)
3 A/m(Ed3.0)
30 A/m(Ed4.0)
3 A/m(Ed3.0)
30 A/m(Ed4.0)
Power frequency magnetic
fields are at levels
characteristic of a typical
location in a typical commercial
and/or hospital environment.
NOTE
:
U
T
is the a.c. mains voltage prior to application of the test level.
Содержание Brivo OEC 715
Страница 9: ...Chapter1 Introduction and Safety...
Страница 34: ...Chapter2 System Overview...
Страница 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Страница 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Страница 40: ...Chapter3 Mechanical Positioning...
Страница 52: ...Chapter4 Start Up and Operating Control...
Страница 77: ...Chapter5 Usual Operations...
Страница 108: ...Chapter6 SpecialApplications...
Страница 139: ...Chapter7 Radiographic Film...
Страница 142: ...Chapter8 System Setup...
Страница 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Страница 150: ...Chapter9 Maintenance...
Страница 160: ...Chapter10 Display Messages...
Страница 162: ...Chapter11 Labels and Symbols...
Страница 167: ...Labels and Symbols Page 11 6 18 1 System rating plate indicates manufacture information and input power requirements...
Страница 168: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 11 7 19 1 X ray Generator rating plate Total Filtration label...
Страница 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Страница 179: ...Chapter12 Technical Reference...
Страница 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Страница 194: ...Technical Reference Page 12 16 Lateral configuration 1 5m LOHMANN Lateral configuration 1m LOHMANN...
Страница 206: ...Technical Reference Page 12 28 12 11 System Block...
Страница 209: ......
Страница 210: ...www gehealthcare com Imagination at work...