Technical Reference
Page 12-24
CAUTION
6.
After attaching the laser aimer, place the C-Arm in the position it will be used during the procedure.
7.
Turn the laser aimer ON and adjust the beam alignment knobs until the laser beam is centered on the
crossing of the cross-hair on the tube cover.
If the laser is ON, do not look directly into the laser
aimer’s aperture.
8.
Using fluoroscopy, position the cross-hair in the fluoro image, at the center of the area of interest.
The location where the laser beam marks the patient corresponds to the center of the area
of interest, indicated by the cross-hair in the fluoro image.
If you
don’t want the cross hair to show on the image, remove the laser cover from the laser aimer.
1.
Switch OFF the power to the laser aimer.
2.
Grasp the two knobs on the laser cover.
3.
Press down and rotate the cover (approximately 10 degrees counterclockwise) until the knobs align
with the notches on the housing.
4.
Lift off the cover and resume laser aimer operation.
Regardless of this procedure to adjust or operate the equipment may result harmful
radiation.
12.7.3.
30cm Skin Spacer
Skin Spacer is able to install on the tube cover and removable filter. Install the skin spacer as the following
steps:
1.
Put the skin spacer on the X-Ray tube cover or removable filter. Make sure the s
kin spacer’s hole
superpose with the screws.
2.
Rotate the skin spacer clockwise to lock it onto the cover.
NOTE
WARNING
Содержание Brivo OEC 715
Страница 9: ...Chapter1 Introduction and Safety...
Страница 34: ...Chapter2 System Overview...
Страница 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Страница 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Страница 40: ...Chapter3 Mechanical Positioning...
Страница 52: ...Chapter4 Start Up and Operating Control...
Страница 77: ...Chapter5 Usual Operations...
Страница 108: ...Chapter6 SpecialApplications...
Страница 139: ...Chapter7 Radiographic Film...
Страница 142: ...Chapter8 System Setup...
Страница 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Страница 150: ...Chapter9 Maintenance...
Страница 160: ...Chapter10 Display Messages...
Страница 162: ...Chapter11 Labels and Symbols...
Страница 167: ...Labels and Symbols Page 11 6 18 1 System rating plate indicates manufacture information and input power requirements...
Страница 168: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 11 7 19 1 X ray Generator rating plate Total Filtration label...
Страница 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Страница 179: ...Chapter12 Technical Reference...
Страница 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Страница 194: ...Technical Reference Page 12 16 Lateral configuration 1 5m LOHMANN Lateral configuration 1m LOHMANN...
Страница 206: ...Technical Reference Page 12 28 12 11 System Block...
Страница 209: ......
Страница 210: ...www gehealthcare com Imagination at work...