![GE Brivo OEC 715 Скачать руководство пользователя страница 67](http://html1.mh-extra.com/html/ge/brivo-oec-715/brivo-oec-715_operators-manual_4193902067.webp)
Start Up and Operating Control
Page 4-16
WARNING
4.3.3.
Emergency Switch
The emergency switch located on the top of C-Arm control panel. When emergency situation occurs, press
the switch to stop vertical column mechanical movement and disable X-Rays.
If pressing an emergency switch fails to stop motor movement or X-Rays, press the power
off switch or disconnect the power plug from the AC receptacle.
If the emergency switch is pressed on purpose or inadvertently, vertical column operation will cease and
X-Rays will be disabled. Rotate the Emergency Switch in the direction of the arrow indication to release it
and continue to use the system.
Should the system unexpectedly become unresponsive or lockup, restart the system to return it to normal
operating condition. If the system fails to respond or continues to lockup, call your service representative.
Emergency Switch
Содержание Brivo OEC 715
Страница 9: ...Chapter1 Introduction and Safety...
Страница 34: ...Chapter2 System Overview...
Страница 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Страница 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Страница 40: ...Chapter3 Mechanical Positioning...
Страница 52: ...Chapter4 Start Up and Operating Control...
Страница 77: ...Chapter5 Usual Operations...
Страница 108: ...Chapter6 SpecialApplications...
Страница 139: ...Chapter7 Radiographic Film...
Страница 142: ...Chapter8 System Setup...
Страница 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Страница 150: ...Chapter9 Maintenance...
Страница 160: ...Chapter10 Display Messages...
Страница 162: ...Chapter11 Labels and Symbols...
Страница 167: ...Labels and Symbols Page 11 6 18 1 System rating plate indicates manufacture information and input power requirements...
Страница 168: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 11 7 19 1 X ray Generator rating plate Total Filtration label...
Страница 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Страница 179: ...Chapter12 Technical Reference...
Страница 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Страница 194: ...Technical Reference Page 12 16 Lateral configuration 1 5m LOHMANN Lateral configuration 1m LOHMANN...
Страница 206: ...Technical Reference Page 12 28 12 11 System Block...
Страница 209: ......
Страница 210: ...www gehealthcare com Imagination at work...