Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 4-25
NOTE
Is the
“
Edge Enhancement
”
level
set too low?
Increase edge enhancement
level setting.
Is the Image Intensifier far away
from the patient?
Re-position the Image Intensifier
closer to patient.
Is the region of interest not at the
center of the image?
Re-position the region of interest
to the center of the image field.
Is the
“
Field Size
”
set to NORM?
Switch to MAG1/MAG2 mode.
7
Small field of view
(images look too big)
Is the
“
Field Size
”
set to
MAG1/MAG2 mode?
Switch to NORM field size.
Is the Image Intensifier far away
from the patient?
Re-position the Image Intensifier
closer to Patient.
Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product also provides the feature of Auto-Adaptive
Non-uniformity Correction to improve the uniformity of the image. Because of the curved nature of the
image intensifier tube, less X-Ray energy is received at the periphery of the tube than at the center. This
causes the image to appear darker around the image edge. Auto-Adaptive Non-uniformity Correction is
used to correct it and improve the uniformity of the image.
Correction template can be created automatically in service mode by service engineer, and
then the images in live mode can be corrected with the template in real time. This feature is
not user selectable.
Содержание Brivo OEC 715
Страница 9: ...Chapter1 Introduction and Safety...
Страница 34: ...Chapter2 System Overview...
Страница 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Страница 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Страница 40: ...Chapter3 Mechanical Positioning...
Страница 52: ...Chapter4 Start Up and Operating Control...
Страница 77: ...Chapter5 Usual Operations...
Страница 108: ...Chapter6 SpecialApplications...
Страница 139: ...Chapter7 Radiographic Film...
Страница 142: ...Chapter8 System Setup...
Страница 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Страница 150: ...Chapter9 Maintenance...
Страница 160: ...Chapter10 Display Messages...
Страница 162: ...Chapter11 Labels and Symbols...
Страница 167: ...Labels and Symbols Page 11 6 18 1 System rating plate indicates manufacture information and input power requirements...
Страница 168: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 11 7 19 1 X ray Generator rating plate Total Filtration label...
Страница 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Страница 179: ...Chapter12 Technical Reference...
Страница 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Страница 194: ...Technical Reference Page 12 16 Lateral configuration 1 5m LOHMANN Lateral configuration 1m LOHMANN...
Страница 206: ...Technical Reference Page 12 28 12 11 System Block...
Страница 209: ......
Страница 210: ...www gehealthcare com Imagination at work...