Action
Step
Place the cassette tray on the tray support of the fraction collector drawer.
Make sure that the front of the tray (marked with the GE monogram) faces
the front of the drawer and is hooked onto the two pins.
6
Close the drawer. Make sure that it snaps into closed position.
7
Result: After the door has been closed, the fraction collector performs a Full
scan of the cassette type code of each cassette to identify the cassette
types, see
Cassette and tray identification, on page 127
.
Prepare a rack for 50 ml tubes or
250 ml bottles
Action
Step
Place 50 ml tubes or 250 ml bottles in the corresponding rack (rack with 50
ml tubes shown in illustration).
1
ÄKTA avant Operating Instructions 29101556 AB
125
5 Prepare the system for a run
5.7 Prepare the built-in fraction collector