Action
Step
Connect a 25 to 30 ml syringe to the outlet tubing of the system pump piston
rinsing system. Draw liquid slowly into the syringe.
6
Disconnect the syringe and discard its contents.
7
Immerse the outlet tubing in the rinsing solution tube where the inlet tubing
of the system pump piston rinsing system is immersed.
8
Connect a 25 to 30 ml syringe to the outlet tubing from the sample pump
piston rinsing system. Draw liquid slowly into the syringe.
9
Disconnect the syringe and discard its contents.
10
Immerse the outlet tubing in the rinsing solution tube where the inlet tubing
of the sample pump piston rinsing system is immersed.
11
Fill the rinsing solution tubes so that each of the tubes contains 50 ml of
20% ethanol.
12
ÄKTA avant Operating Instructions 29101556 AB
79
4 Installation
4.2 Hardware installation
4.2.5 Prepare the pump rinsing system