Allow space on the laboratory bench for:
•
handling of samples and buffers (2 × 30 cm)
•
computer and monitor (80 cm)
•
access for service (see the following topic)
Service access
To access the rear panel, the instrument can be rotated on a swivel foot. There must be
at least 20 cm additional space on the bench to allow for free rotation.
WARNING
Rotating the instrument. Make sure that there is always at least
20 cm of free space around the ÄKTA avant instrument to allow
for sufficient ventilation and rotation on the swivel foot. When ro-
tating the instrument, take care not to stretch or squeeze tubing
or cables. A disconnected cable may cause power interruption or
network interruption. Stretched tubing may cause bottles to fall,
resulting in liquid spillage and shattered glass. Squeezed tubing
may cause increase in pressure, or block liquid flow. To avoid the
risk of knocking over bottles, always place bottles on the buffer
tray, and close the doors before rotating the instrument.
Laboratory bench
The bench must be clean, flat and stable to support the weight of the ÄKTA avant system,
see the following table
.
56
ÄKTA avant Operating Instructions 29101556 AB
4 Installation
4.1 Site preparation
4.1.2 Room requirements