Uninterruptible power supply
(UPS)
A UPS can prevent data loss during a power failure, and allow time for a controlled shut-
down of ÄKTA avant.
For UPS power requirements, see the system specifications in this manual. Remember
to also take into account the specifications for the computer and monitor. Refer to the
manufacturers' documentation.
If using a UPS, the ÄKTA avant instrument, the computer and the monitor must
be connected to the UPS.
Note:
Restart the instrument after
emergency shutdown or power
failure
Follow the instructions to restart the instrument after an emergency shutdown or power
failure.
Action
Step
Make sure that the condition that caused the emergency shutdown or
power failure is corrected.
1
If the instrument was switched off, press the Power switch on the instrument.
2
Result: The instrument should start and the Instrument display should show
Not connected.
Turn on the computer and monitor.
3
Start UNICORN and connect to the system.
4
See instructions in
Section 4.4 Start UNICORN and connect to system, on
32
ÄKTA avant Operating Instructions 29101556 AB
2 Safety Instructions
2.3 Emergency procedures