General Alarms and Troubleshooting
COBE
®
Spectra™ Apheresis System •
Essentials Guide
12-3
If a “RETURN PRESSURE HIGH! Decrease return flow scale” alarm
occurs near the end of a Draw phase, and a “RETURN FLOW TOO
SLOW! Check return. Increase return flow scale.” alarm occurs near the
end of the next Return phase, reduce the inlet pump flow rate. This
alarm can also occur if you do not completely turn the Return Flow
Controller clockwise during Prime mode.
These conflicting alarm messages can occur when the Single-Needle
pump flow rates are too high to support a specific combination of return
flow restrictions (Single-Needle, needle position, high return hematocrit,
high return pressure, or a restricted return path) and the 400 mm Hg
return pressure limit.
Alarm Clearing
If you cannot clear an alarm by following the instructions in this chapter,
try turning the Spectra system’s power switch off and on and then
continuing the procedure.
If the alarm occurs during Prime mode, follow these steps:
1
Press
CHANGE MODE
.
2
Press
1
to select “Load Set.”
3
Select the correct disposable tubing set and reprime.
Warnings and Alarms
The following table lists all the warnings and alarms, and the page on
which you can find the detailed information about each.
Note:
If a “RETURN FLOW TOO SLOW!” alarm occurs, always
check for the possibility of an infiltration, a blockage at the
donor/patient connection site, or an improperly positioned return
pressure sensor transducer.
Table 12-1:
Warnings and Alarms
Warning or Alarm Message
See Page…
*
AC infusion rate configuration was changed last run..
12-8
AC infusion rate exceeds (allowable) limit(s)..
12-9
AC PUMP ERROR!
12-10
ACCESS PRESSURE ERROR!
12-11
ACCESS PRESSURE LOW.
12-12
ACCESS PRESSURE LOW! Check access line and needle..
12-13
ACCESS PRESSURE OCCLUSION ERROR!
12-14
ACCESS PRESSURE SENSOR DID NOT REACH ALARM LIMIT!
12-15
ACCESS PRESSURE SENSOR NOT ZERO!
12-16
AIR IN INLET CHAMBER!
12-17
AIR IN RETURN CHAMBER!
12-18
ALARM TEST FAILED!
12-19
Assisted Mode
4-25,
CT Guide
CCM CALIBRATION FAILURE..
12-19
CCM INDICATES NO PLATELETS COLLECTED.
6-2,
PC Guide
CCM not operational - this run only..
12-20
Содержание Cobe Spectra
Страница 1: ......
Страница 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Страница 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Страница 51: ...COBE Spectra Apheresis System Essentials Guide 3 1 3 Installing Transporting and Returning the Spectra System ...
Страница 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Страница 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Страница 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Страница 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Страница 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Страница 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Страница 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Страница 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Страница 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Страница 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Страница 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Страница 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Страница 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Страница 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...