Introduction
2-10
COBE
®
Spectra™ Apheresis System •
Essentials Guide
PLASMA
Plasma volume is a limiting factor for TPE procedures. You can concurrently collect plasma with platelets and WBCs.
•
During the Run mode, the current accumulated volume of plasma (and anticoagulant in mL) collected appears
above the PLASMA key.
•
When the target values appear, press
PLASMA
to change the target plasma volume (in mL).
COLLECT/REPLACE
Collect volume is a limiting factor for ELP, AutoPBSC, LRS, LRST, and WBC procedures. Replacement volume is a
limiting factor for TPE and RBCX procedures.
•
During the Run mode, the current accumulated volume appears above the COLLECT/REPLACE key. This volume
represents either:
•
Amount collected (mL) for a donor collection or patient depletion procedure.
•
Amount of replacement fluid (mL) used during a TPE or RBCX procedure.
•
When the target values appear, press
COLLECT/REPLACE
to change the target collect volume (mL) during ELP,
AutoPBSC, LRS, and WBC procedures. To change the target replacement volume (mL) during TPE and RBCX
procedures, press
COLLECT/REPLACE
.
Table 2-4:
PROCEDURE key functions
PROCEDURE keys
Function
TIME MIN
Run time is a limiting factor for some procedures.
•
During the Run mode, the elapsed procedure time (min) appears above the TIME MIN key.
•
To change the target run time, when the target end results appear press
TIME MIN
.
This key does not function during RBCX procedures.
TARGET VALUES
Press
TARGET VALUES
to toggle back and forth between target values and the current values on the bottom line of
the display. When the system displays target values, “Target” appears in the bottom right corner of the screen. With
the exception of the AC volume, to change the target values press
TARGET VALUES
, press the appropriate VOLUME
key below the display, and enter the new target value.
When the current values appear, a procedure abbreviation appears in the bottom right corner. See the list on page
2-14 for the apheresis procedure abbreviations.
Table 2-3:
VOLUME key functions (continued)
VOLUME keys
Function
Содержание Cobe Spectra
Страница 1: ......
Страница 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Страница 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Страница 51: ...COBE Spectra Apheresis System Essentials Guide 3 1 3 Installing Transporting and Returning the Spectra System ...
Страница 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Страница 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Страница 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Страница 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Страница 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Страница 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Страница 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Страница 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Страница 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Страница 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Страница 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Страница 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Страница 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Страница 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...