General Alarms and Troubleshooting
12-48
COBE
®
Spectra™ Apheresis System •
Essentials Guide
PLASMA VALVE NOT OPERATING CORRECTLY!
Alarm LED is red. Pumps stop.
To retry the plasma valve test, press
CONTINUE
.
This alarm occurs only during TPE and RBCX procedures when the
plasma pump fails to generate expected pressures at the return pressure
sensor for various plasma valve positions.
PLASMA VALVE NOT OPERATING CORRECTLY!
Check for proper loading.
CONTINUE
Probable Cause
Action
Tubing incorrectly positioned in
plasma valve
Ensure the tubing is placed correctly in the plasma valve. Press
CONTINUE
.
Replacement fluid or collection bag
lines clamped
Ensure the replacement or collection bag lines are unclamped. Press
CONTINUE
.
Leak at the return pressure sensor
pod seal
Check the return pressure sensor pod seal for wear or damage. Replace if necessary. For instructions, see “Changing pressure pod
seals” on page 10-4.
Note:
For instructions on repositioning the diaphragm and reloading the return pressure sensor, see “Repositioning the
Return Pressure Sensor Diaphragm” on page 11-10.
Return Flow Controller incorrectly set
Ensure the Return Flow Controller is completely in the prime position; or during Run mode, it is in the proper setting.
Plasma valve in the wrong position
Note the alarm message. Notify a qualified service technician of the alarm message.
If alarm occurs during the Run mode of a procedure, use a flat-blade screwdriver to manually place the valve in the proper position.
Pump malfunction
Note the alarm message. Discontinue the procedure. Notify a qualified service technician of the alarm message.
Содержание Cobe Spectra
Страница 1: ......
Страница 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Страница 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Страница 51: ...COBE Spectra Apheresis System Essentials Guide 3 1 3 Installing Transporting and Returning the Spectra System ...
Страница 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Страница 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Страница 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Страница 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Страница 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Страница 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Страница 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Страница 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Страница 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Страница 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Страница 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Страница 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Страница 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Страница 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...