General Alarms and Troubleshooting
COBE
®
Spectra™ Apheresis System •
Essentials Guide
12-47
PLASMA PUMP ERROR!
Alarm LED is red. Pumps stop.
To retry the plasma pump test, press
CONTINUE
.
This alarm occurs only during Prime mode if the plasma pump fails to
generate an expected pressure increase at the return pressure sensor.
.
PLASMA PUMP ERROR! Check for proper
loading or leaks.
CONTINUE
Probable Cause
Action
Return pressure sensor not correctly loaded in the sensor
housing
Ensure the return pressure sensor is correctly loaded in the return pressure sensor housing. See
“Repositioning the Return Pressure Sensor Diaphragm” on page 11-10. Press
CONTINUE
.
Plasma, return, or RBC valve incorrectly loaded
Ensure the valves are loaded correctly. Press
CONTINUE
.
Male/female luer not connected
Connect the luer according to the disposable tubing set in use.
Incorrect disposable tubing set selected during data entry
Press
CHANGE MODE
, press
1
to select “Load Set,” choose the correct disposable tubing set, reprime the set.
Plasma line occluded
Ensure the plasma line is not kinked, twisted, or pinched, and that there is fluid in the line. Press
CONTINUE
.
Plasma line inadvertently placed in the RBC valve housing
Remove the plasma line from the RBC valve housing and place the RBC line in the RBC valve. Press
CONTINUE
.
Leak in disposable tubing set
Check the disposable tubing set for leaks. If the set does not leak, press
CONTINUE
. If the set leaks, replace it
with a new set and prime the set.
If the alarm recurs, replace the disposable tubing set.
Return Flow Controller incorrectly set
Ensure the Return Flow Controller is completely in prime position.
Leak at the return pressure sensor pod seal
Check the return pressure sensor pod seal for wear or damage. Replace if necessary. For instructions, see
“Changing pressure pod seals” on page 10-4.
Note:
For instructions on repositioning the diaphragm and reloading the return pressure sensor,
see “Repositioning the Return Pressure Sensor Diaphragm” on page 11-10.
Pump, valve, or sensor malfunction
Note the alarm message. Discontinue the procedure. Notify a qualified service technician of the alarm
message.
Содержание Cobe Spectra
Страница 1: ......
Страница 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Страница 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Страница 51: ...COBE Spectra Apheresis System Essentials Guide 3 1 3 Installing Transporting and Returning the Spectra System ...
Страница 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Страница 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Страница 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Страница 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Страница 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Страница 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Страница 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Страница 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Страница 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Страница 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Страница 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Страница 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Страница 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Страница 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...