General Alarms and Troubleshooting
12-42
COBE
®
Spectra™ Apheresis System •
Essentials Guide
NO SALINE SEEN AT INLET AIR SENSOR!
Alarm LED is red. Pumps stop.
Press
CONTINUE
to retry the priming inlet test.
This alarm occurs only during Prime mode if the inlet air sensor still
detects air after a specific time period.
NO SALINE SEEN AT INLET AIR SENSOR!
Check sensor for loading.
CONTINUE
Probable Cause
Action
Inlet air chamber not placed in the inlet air detector
Ensure the inlet air chamber is placed correctly in the inlet air detector. Press
CONTINUE
.
Saline container not connected
– or –
Access saline line clamp not open
Ensure the saline container is connected and that the access saline line roller clamp is open. Press
CONTINUE
.
Sensors not detecting fluid in inlet air chamber
Reposition the inlet air chamber. Ensure it has good contact with the sensor and the filter is below the sensor.
Press
CONTINUE
.
Centrifuge pressure sensor tubing out of place
Reinsert the centrifuge pressure sensor tubing. To reprime the disposable tubing set, press
CHANGE MODE
,
then press
2
to select “Prime.”
Return line valve unloaded
Reload the return line valve. Press
CONTINUE
.
Plasma line inserted in RBC valve
Ensure the RBC line is in the RBC valve. Press
CONTINUE
.
Pump or sensor malfunction
Note the alarm message. Discontinue the procedure. Notify a qualified service technician of the alarm
message.
Содержание Cobe Spectra
Страница 1: ......
Страница 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Страница 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Страница 51: ...COBE Spectra Apheresis System Essentials Guide 3 1 3 Installing Transporting and Returning the Spectra System ...
Страница 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Страница 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Страница 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Страница 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Страница 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Страница 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Страница 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Страница 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Страница 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Страница 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Страница 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Страница 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Страница 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Страница 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...