General Alarms and Troubleshooting
12-20
COBE
®
Spectra™ Apheresis System •
Essentials Guide
CCM not operational - this run only.
Warning and Alarm LEDs are off.
This message appears only if the CCM display is selected during the Run
mode and saline calibration (during Prime mode) was not successful. A
CCM CALIBRATION FAILURE alarm should have occurred during
Prime mode.
CCM not operational – this run only.
Probable Cause
Action
Saline calibration failed at end of Prime mode due to a
problem with cuvette installation (cuvette contained bubbles
or it was installed incorrectly)
To remove the message, press
CLEAR
. The CCM is not accurate during this Run mode only. Monitor the
channel for RBC accumulation. During AutoPBSC procedures, open the collect valve during the Harvest
phase (see “Assisted Mode” in Chapter 4 of the
Cell Therapy Guide
).
CCM malfunction
If the message continues to appear, notify a qualified service technician.
Содержание Cobe Spectra
Страница 1: ......
Страница 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Страница 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Страница 51: ...COBE Spectra Apheresis System Essentials Guide 3 1 3 Installing Transporting and Returning the Spectra System ...
Страница 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Страница 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Страница 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Страница 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Страница 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Страница 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Страница 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Страница 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Страница 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Страница 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Страница 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Страница 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Страница 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Страница 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...