Introduction
2-24
COBE
®
Spectra™ Apheresis System •
Essentials Guide
Unload Set
Once Rinseback mode is complete and the donor/patient is
disconnected, the system displays the final run values and prompts you
to unload the pumps. The inlet pump rotates to verify that the
donor/patient is disconnected. After successful completion of the donor
disconnect test, the pumps unload.
.
Diagnostics
After it loads and primes the tubing set, the system gives you the option
to complete alarm tests. See Chapter 7, “Alarm Tests,” for more
information.
Automatic and Manual Mode
The Spectra system operates in Automatic or Manual mode. The system
normally operates in Automatic mode, but can operate in Manual mode
for certain recovery procedures. For additional information on
Automatic and Manual modes, see Chapter 6, “Automatic and Manual
Modes.”
Anticoagulation
Anticoagulation is necessary to prevent clotting in the extracorporeal
circuit. It also establishes a pH and ionized calcium environment that
prevents cell clumping. For a thorough discussion of anticoagulation
during Spectra system procedures, see Chapter 8, “Anticoagulation.”
Note:
Once the pumps unload, the final run values are no longer
available.
Содержание Cobe Spectra
Страница 1: ......
Страница 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Страница 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Страница 51: ...COBE Spectra Apheresis System Essentials Guide 3 1 3 Installing Transporting and Returning the Spectra System ...
Страница 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Страница 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Страница 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Страница 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Страница 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Страница 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Страница 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Страница 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Страница 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Страница 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Страница 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Страница 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Страница 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Страница 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...