Loading and Removing the Disposable Tubing Set and Other Common Operations
4-14
COBE
®
Spectra™ Apheresis System •
Essentials Guide
The Spectra version 7.0 screen:
Label the product bags to notify the laboratory that this product is
leukoreduced.
– or –
If the platelet yields could be less than predicted and/or events
occurred during the procedure that may cause WBC
contamination, one of the following screens appears:
The Spectra version 5.1 screen:
The Spectra version 7.0 screen:
Label the product to notify the laboratory to measure this product
for platelet and WBC content.
The second screen above appears in LRST procedures. See
“Measuring Platelet and WBC Content Codes” in Chapter 5 of the
Platelet Collection Guide
for the codes and their meanings.
WBC content is usually below 1 x 10
6
, but actual WBC content
should be verified to ensure the product meets local standards for
leukoreduction.
6
The following screen appears:
ELP, LRS, and LRST procedures
:
IMPORTANT
:
Seal the collect
line above the connection and remove the collection bags. For
instructions on collecting a multiple platelet product, see “Preparing
a Multiple Platelet Product” in Chapter 5 of the
Platelet Collection
Guide
.
If you concurrently collected plasma, seal the plasma line and
remove the plasma product bag.
Platelet Depletion procedures: IMPORTANT
:
Close the slide
clamps on the line to the platelet bags.
MNC, PMN, and AutoPBSC procedures
:
IMPORTANT
:
Seal the
collect line and remove the collection and plasma bags.
WBC Depletion and LPE procedures:
IMPORTANT
:
Clamp or
seal the collect line and remove the collection bag(s).
TPE procedures:
IMPORTANT
:
Close the clamp on the line
leading to the plasma bag and replacement fluid bags.
Dual-Needle procedures:
Disconnect the access needle and place it
in an appropriate needle disposal container, or disconnect the access
line from the vascular access device.
7
Press
CLEAR
.
During Rinseback, the Spectra system displays several operation screens,
and pumps saline to the channel to flush RBCs to the donor/patient.
Disconnecting the Donor or Patient
Follow the instructions in the procedural guides.
Label LRS Turbo product <1E6 WBC
content. Press
CLEAR
Measure platelet and WBC content of
this product!
Press CLEAR
Measure platelet and WBC content!
(CODE)
Press CLEAR
Clamp and disconnect collection bags
Press CLEAR
Содержание Cobe Spectra
Страница 1: ......
Страница 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Страница 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Страница 51: ...COBE Spectra Apheresis System Essentials Guide 3 1 3 Installing Transporting and Returning the Spectra System ...
Страница 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Страница 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Страница 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Страница 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Страница 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Страница 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Страница 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Страница 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Страница 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Страница 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Страница 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Страница 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Страница 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Страница 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...