63
Section 10
Cleaning Optical Ports
The Voice, Test, and OPM Ports must be kept free from dirt or other contaminates to ensure accurate
measurements and operation. A supply of optical cleaning pads and isopropyl alcohol or a connector-
cleaning cartridge is recommended to clean the optical ports. Before conducting the cleaning procedures
that follow, turn the instrument off.
Cleaning Voice and Test Ports
1 Using the supplied hex key screw driver, remove the two hex head screws on the Removable Cleaning
Port.
2 Pull the Removable Cleaning Port directly out away from the mount.
3 The ferrule can now be cleaned. Perform one of the following:
If using Noyes Cleaning Supplies
• Leaning the FCC2 can back (30°), press the button on the FCC2 to fill the well.
• Dip the CCTP tip into the well of the FCC2 to dampen the tip with optical cleaning fluid.
• Place the damp tip over the ferrule to be cleaned.
• Rotate the tip clockwise 10 revolutions while applying varying pressure to create a gentle pumping
action where the tip contacts the ferrule.
• Discard the CCTP stick after using both tips.
Figure 10-1. Cleaning Optical Ports
Voice
Test Port
OPM
Removable Cleaning Port
2
Internal
Ferrule
7
Adapter Cap
2
1
1
Hex Screw
Hex Screw