37
Section 5
Figure 5-4. Link Test Setup.
Performing a Link Test
Refer to Figure 5-4 for a related illustration.
1 Attach the APC connector on the reference cable to the Test Port.
2 Attach the UPC connector on the reference cable to the fiber under test at the patch panel.
Note:
If the link under test is less than 60 km, then the use of a Terminator is recommended. If the
link is greater than 60 km, then termination is not required.
3 Once connected, the TURBOTEST 500B will automatically measure and display the return loss
measurements for the link under test.
4 Press the [SAVE] soft key to store test results in the next available memory location.
5 To save test results in a specific memory location, perform the following steps:
• Press either [UP] or [DOWN] arrow key to select a specific memory location (increase/decrease
fiber #).
• Press the [SAVE] soft key to store test results in the selected location.
Once saved, test results can later be retrieved. (refer to section titled “Reviewing Saved ORL Test Results”, page 38).
Terminator
Patch Panel
UPC Connector
Link under test
TURBOTEST
500B
APC Connector
Reference Cable
Test Port
Patch Panel