A-2
Appendix
A Specifications
DR-ID1270 Operation Manual 897N120821E
■ Battery charger (optional)
(1) Operating Conditions
Temperature
: 0°C - 35°C
Humidity
: 10%RH - 85%RH (no dew condensation)
(2) Non-operating Conditions
(Environmental conditions under which power can be supplied)
Temperature
: -20°C - 60°C
Humidity
: 10%RH - 95%RH (no dew condensation)
CAUTIONS
Charge the battery pack in the operating environment.
A.1.4 Image Performance
Each flat panel sensor complies with IEC 62220-1 (MEDICAL ELECTRICAL EQUIPMENT -
CHARACTERISTICS OF DIGITAL X-RAY IMAGING DEVICES - ) as a general X-ray radiography
equipment.
To ensure optimal image quality, it is recommended that you do not use the flat panel sensor near
devices (motor, transformer, switching supply, etc.) that generate electromagnetic noise.
A.1.5 Load Restriction
Entire surface load :
DR-ID1271SE, DR-ID1272SE, DR-ID1273SE, DR-ID1274SE, DR-ID1275SE :
300kg (661.5 lb)
Local load :
DR-ID1271SE, DR-ID1272SE, DR-ID1273SE, DR-ID1274SE, DR-ID1275SE :
120kg (264.6 lb) / ø40mm (1.6in.) (Based on FUJIFILM measurement specifications)
CAUTIONS
Do not apply an excessive force to the exposure plane.
The sensor inside the flat panel sensor may be damaged, and it may not be possible to make
an exposure properly.
A.1.6 Radio Waves
Wireless specifications for the flat panel sensor are as follows.
Flat panel sensor
Wireless specification
IEEE802.11n
Transmit frequency
5.2, 5.3, 5.6, 5.8, 2.4 GHz
Modulation
OFDM
Frequency tolerance
±20 ppm
Data transfer rate
35 Mbps
Transfer power
17 dBm or less
CAUTIONS
● Transmit frequencies available vary, depending on the country.
● Radio waves available outdoors vary, depending on the country where the system is used.
● When the DR-ID1270 and any other wireless equipment are operating on the same frequency
channel in a hospital, it may take time to show an image on the image processing unit
monitor.
Содержание FDR ES DR-ID1270
Страница 2: ...ii DR ID1270 Operation Manual 897N120821E...
Страница 28: ...1 22 For Safe Operation 1 DR ID1270 Operation Manual 897N120821E...
Страница 42: ...2 14 System Configuration Product Overview 2 DR ID1270 Operation Manual 897N120821E...
Страница 54: ...3 12 Basic Operation 3 DR ID1270 Operation Manual 897N120821E...
Страница 56: ...4 2 Daily Inspection and Maintenance 4 DR ID1270 Operation Manual 897N120821E...
Страница 68: ...Z 6 Appendix Z Precautions for Exposure DR ID1270 Operation Manual 897N120821E...
Страница 74: ...O 6 Appendix O Use of Optional Items DR ID1270 Operation Manual 897N120821E...
Страница 76: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...