Maintenance and Inspection
DR-ID1270 Operation Manual 897N120821E
Maintenance and Inspection
1 Maintenance and Inspection Items Assigned to Specified Dealer
For periodical inspection of the equipment and necessary arrangements, consult our official dealer
or local representative.
Periodical Maintenance
Make sure that the periodical maintenance and inspection assigned to our official dealer or
FUJIFILM Representative are performed as specified.
Maintenance and Inspection Items Assigned to Specified Dealer
Periodical Maintenance and Inspection Items
Period
Checking of the image
Every year
Checking of the operation record by referring to the error log
Every year
Checking of the units
Every 2 years
Main Periodical Replacement Parts
Name of Periodical Replacement Parts
Period
Relay (optional)
Every 1 years
(Number of exposures : 90,000)
* It is recommended that the battery pack be replaced, if the battery storage capacity becomes
lower than 60%.
The battery pack should be replaced when the operable time is less than the following.
●
DR-ID1271SE/DR-ID1272SE/DR-ID1273SE/DR-ID1274SE/DR-ID1275SE :
100 minutes (1 hour and 40 minutes)
* Refer to the operable time displayed on the image processing unit when the battery pack is fully
charged and no exposure menu is registered.
* Depending on the usage environment, etc., the displayed time is slightly different from the
actual operable time.
(Reference) Time for battery pack replacement
Battery storage capacity [%]
Operable time [Minute]
120
100
80
60
60
40
20
0
0
30
60
90
120
150
210
240
180
Time for replacement:
Less than 100 minutes
(1 hour and 40 minutes)
The cycles of periodical maintenance and inspection and of parts replacement differ depending on
the usage and the daily operation time.
For details, contact our official dealer or FUJIFILM Representative.
Содержание FDR ES DR-ID1270
Страница 2: ...ii DR ID1270 Operation Manual 897N120821E...
Страница 28: ...1 22 For Safe Operation 1 DR ID1270 Operation Manual 897N120821E...
Страница 42: ...2 14 System Configuration Product Overview 2 DR ID1270 Operation Manual 897N120821E...
Страница 54: ...3 12 Basic Operation 3 DR ID1270 Operation Manual 897N120821E...
Страница 56: ...4 2 Daily Inspection and Maintenance 4 DR ID1270 Operation Manual 897N120821E...
Страница 68: ...Z 6 Appendix Z Precautions for Exposure DR ID1270 Operation Manual 897N120821E...
Страница 74: ...O 6 Appendix O Use of Optional Items DR ID1270 Operation Manual 897N120821E...
Страница 76: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...