A-1
Appendix
A Specifications
DR-ID1270 Operation Manual 897N120821E
Appendix A
Specifications
A.1 Specifications
Specifications of the DR-ID1270 are shown below.
A.1.1 Reduced Equivalent (DR-ID1270)
Peak reduced equivalent on the front panel of the flat panel sensor: 0.5 mmAl
A.1.2 Power Supply Conditions
■ Flat panel sensor
Rated voltage : 6-12V
Input current : 2.73A
■ Power supply unit/Power box
Rated voltage : 100-240V~
Input current : 2-0.84A
Frequency : 50-60Hz
A.1.3 Environmental Conditions
■
Flat panel sensor
/DR-ID1270MP/DR-ID1270PB
(1) Operating Conditions
Temperature
: 15°C (15%RH) - 35°C (80%RH)
Humidity
: 15%RH (15°C) - 80%RH (30°C) (no dew condensation)
Atmospheric pressure : 700hPa - 1060hPa
(2) Non-operating Conditions
(Environmental conditions under which power can be supplied)
Temperature
: 5°C - 35°C
Humidity
: 10%RH - 80%RH (no dew condensation)
Atmospheric pressure : 700hPa - 1060hPa
(3) Storage Conditions
Temperature
: -30°C - 50°C
Humidity
: 10%RH - 90%RH (no dew condensation)
Atmospheric pressure : 700hPa - 1060hPa
CAUTIONS
● When the flat panel sensor is used in high temperature condition for long period of time, it
may cause image artifacts and/or failure of the device.
● When using the DR-ID 1271SE, DR-ID 1273SE or DR-ID 1275SE, if the temperature is 37ºC
and the humidity is 90% RH (no dew condensation), continuous use of 30 minutes or less is
possible.
Using Manual Mode (energy saving mode) from the image processing unit when the
temperature and humidity are the same allows up to 1 hour of continuous use.
Содержание FDR ES DR-ID1270
Страница 2: ...ii DR ID1270 Operation Manual 897N120821E...
Страница 28: ...1 22 For Safe Operation 1 DR ID1270 Operation Manual 897N120821E...
Страница 42: ...2 14 System Configuration Product Overview 2 DR ID1270 Operation Manual 897N120821E...
Страница 54: ...3 12 Basic Operation 3 DR ID1270 Operation Manual 897N120821E...
Страница 56: ...4 2 Daily Inspection and Maintenance 4 DR ID1270 Operation Manual 897N120821E...
Страница 68: ...Z 6 Appendix Z Precautions for Exposure DR ID1270 Operation Manual 897N120821E...
Страница 74: ...O 6 Appendix O Use of Optional Items DR ID1270 Operation Manual 897N120821E...
Страница 76: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...