2-1
System Configuration (Product Overview)
2
DR-ID1270 Operation Manual 897N120821E
Chapter 2
System Configuration
(Product Overview)
2.1 DR-ID1270
2.1.1 System Confi guration
DIGITAL RADIOGRAPHY DR-ID1270
Flat panel sensor unit
Flat panel sensor unit
Battery charger
Power supply unit*
1
Flat panel
sensor
DR-ID 1200MC
(software)
*
2
Access point
Image processing unit
DR-ID 300CL*
3
Power box*
1
Flat panel
sensor
• The products in
can be installed in patient environment.
• Up to fi ve fl at panel sensors can be connected. Up to two fl at panel sensors can be connected to
one power supply unit in wired communication mode. Up to one power supply units can be used.
• The DR-ID1270 consists of the fl at panel sensor DR-ID1271SE, DR-ID1272SE, DR-ID1273SE,
DR-ID1274SE, DR-ID1275SE and DR-ID 1200MC.
• When the power supply unit is used, the system can be operated in conjunction with the X-ray
equipment, and it can be used for exposures in wired mode.
• When the Power supply unit is used, up to two fl at panel sensors can be connected to one power
supply unit in wired communication mode. Up to two power supply unit can be used. When the
fl at panel sensors are used with three to four different techniques, two power supply units are
required.
• When the Power box is used, only one fl at panel sensor can be connected to one power box in
wired communication mode. Up to three power boxes can be used in one system.
• For the Power supply unit and Power box, a maximum of fi ve devices can be connected.
*1 The power supply unit and the power box are optional products. The system can be confi gured
without using these products.
*2 The DR-ID 1200MC can be installed in the image processing unit DR-ID 300CL.
*3 For detail specifi cation of image processing unit, please refer to “DR-ID 300CL Operation
Manual”.
Содержание FDR ES DR-ID1270
Страница 2: ...ii DR ID1270 Operation Manual 897N120821E...
Страница 28: ...1 22 For Safe Operation 1 DR ID1270 Operation Manual 897N120821E...
Страница 42: ...2 14 System Configuration Product Overview 2 DR ID1270 Operation Manual 897N120821E...
Страница 54: ...3 12 Basic Operation 3 DR ID1270 Operation Manual 897N120821E...
Страница 56: ...4 2 Daily Inspection and Maintenance 4 DR ID1270 Operation Manual 897N120821E...
Страница 68: ...Z 6 Appendix Z Precautions for Exposure DR ID1270 Operation Manual 897N120821E...
Страница 74: ...O 6 Appendix O Use of Optional Items DR ID1270 Operation Manual 897N120821E...
Страница 76: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...