1.3.6 Precautions Related to the Load Applied to the Flat
Panel Sensor
If excessive load is applied to the fl at panel sensor, use it on a fl at fl oor or platform.
When making an exposure for the patient in a wheelchair or adjustable bed or on a stretcher, the
fl at panel sensor may be deformed (slightly warped).
Flat panel sensor
Flat panel sensor
In case that the fl at panel sensor is deformed, make sure that X-ray images are not adversely
affected before continuing the use of the fl at panel sensor.
The precautions below must also be observed when making an exposure.
• Do not have the patient stand on the fl at panel sensor.
• Do not place the hard devices such as spine board on the fl at panel sensor.
Excessive load is applied locally and the fl at panel sensor may be damaged.
Flat panel sensor
Spine board
Even when the fl at panel sensor is used on a fl at fl oor or platform, it may be damaged if the applied
load exceeds the limit.
Содержание FDR ES DR-ID1270
Страница 2: ...ii DR ID1270 Operation Manual 897N120821E...
Страница 28: ...1 22 For Safe Operation 1 DR ID1270 Operation Manual 897N120821E...
Страница 42: ...2 14 System Configuration Product Overview 2 DR ID1270 Operation Manual 897N120821E...
Страница 54: ...3 12 Basic Operation 3 DR ID1270 Operation Manual 897N120821E...
Страница 56: ...4 2 Daily Inspection and Maintenance 4 DR ID1270 Operation Manual 897N120821E...
Страница 68: ...Z 6 Appendix Z Precautions for Exposure DR ID1270 Operation Manual 897N120821E...
Страница 74: ...O 6 Appendix O Use of Optional Items DR ID1270 Operation Manual 897N120821E...
Страница 76: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...