4-1
Daily Inspection and Maintenance
4
DR-ID1270 Operation Manual 897N120821E
Chapter 4
Daily Inspection and
Maintenance
4.1 Daily User Inspection and Maintenance
During maintenance and inspection, strictly observe precautions contained in “Chapter 1 For Safe
Operation” in this manual for you to use the DR-ID1270 under best conditions.
4.1.1 Periodical Inspection
Inspection Every Three Months
Once every three months, remove any dirt or dust accumulated in each part of the equipment using
a vacuum cleaner or air duster, clean each part with a slightly moistened soft cloth and then wipe off
any moisture with a dry cloth.
See “2.2 Unit Names and the Functions” (page 2-3).
CAUTIONS
Be sure to turn off the power before cleaning each part of the device.
No.
Unit
No.
Unit
No.
Unit
1
Flat panel sensor
2
Power supply unit
3
Power supply unit
Air filter (1)
No.
Unit
4
Power box
CAUTIONS
Ensure sufficient space when cleaning the equipment on a table, etc.
Air filter
Clean the air filter on the rear of the power
supply unit with a vacuum cleaner.
Remove the louver while pressing the lever
on the right side, and then clean the air filter
with a vacuum cleaner after detaching it from
the louver.
Air filter
Содержание FDR ES DR-ID1270
Страница 2: ...ii DR ID1270 Operation Manual 897N120821E...
Страница 28: ...1 22 For Safe Operation 1 DR ID1270 Operation Manual 897N120821E...
Страница 42: ...2 14 System Configuration Product Overview 2 DR ID1270 Operation Manual 897N120821E...
Страница 54: ...3 12 Basic Operation 3 DR ID1270 Operation Manual 897N120821E...
Страница 56: ...4 2 Daily Inspection and Maintenance 4 DR ID1270 Operation Manual 897N120821E...
Страница 68: ...Z 6 Appendix Z Precautions for Exposure DR ID1270 Operation Manual 897N120821E...
Страница 74: ...O 6 Appendix O Use of Optional Items DR ID1270 Operation Manual 897N120821E...
Страница 76: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...