Z-2
Appendix Z Precautions for Exposure
DR-ID1270 Operation Manual 897N120821E
4
When an exposure menu is registered and the system is ready for an exposure, the flat panel
sensor enters X-ray detection mode. If an exposure is not made for a period of time while an
exposure menu is registered, the operating time of the flat panel sensor’s battery pack may be
reduced to half. In addition, the battery pack cannot be charged with an exposure menu registered,
even if the flat panel sensor has a wired connection. For these reasons, do not keep the system on
standby, unless you make an exposure.
Z.1.2 Precautions Related to the X-ray Exposure Time
When delivering radiation, do not set the exposure time beyond the maximum limit specified for the
flat panel sensor at the time of installation. Otherwise, vertical artifacts may appear in the image.