3-10
Basic Operation
3
DR-ID1270 Operation Manual 897N120821E
2
Turn off the display as necessary.
3
Make sure that calibration of the flat panel sensor is completed.
When completed, the READY lamp of the flat panel sensor turns off.
4
When the optional access point is used, remove the access point from the image processing unit.
5
Remove the battery pack from the flat panel sensor. Set these battery packs in the battery
charger.
Battery pack
CAUTIONS
If the control cabinet is included in the system, do not turn off the control cabinet with the main
switch. Shutdown operation may not be performed normally.
CAUTIONS
When the system is shut down, image quality adjustment is performed for obtaining optimal
diagnostic images.
Do not disconnect the SE cable or SE communication cable until system shutdown when the
flat panel sensor is used in wired communication mode.
Remove the battery pack after confirming system shutdown when the flat panel sensor is used
in wireless communication mode.
When the Power supply unit and Power box are used, refer to the respective operation manuals.
Содержание FDR ES DR-ID1270
Страница 2: ...ii DR ID1270 Operation Manual 897N120821E...
Страница 28: ...1 22 For Safe Operation 1 DR ID1270 Operation Manual 897N120821E...
Страница 42: ...2 14 System Configuration Product Overview 2 DR ID1270 Operation Manual 897N120821E...
Страница 54: ...3 12 Basic Operation 3 DR ID1270 Operation Manual 897N120821E...
Страница 56: ...4 2 Daily Inspection and Maintenance 4 DR ID1270 Operation Manual 897N120821E...
Страница 68: ...Z 6 Appendix Z Precautions for Exposure DR ID1270 Operation Manual 897N120821E...
Страница 74: ...O 6 Appendix O Use of Optional Items DR ID1270 Operation Manual 897N120821E...
Страница 76: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...