1-12
For Safe Operation
1
DR-ID1270 Operation Manual 897N120821E
Further Information for IEC 60601-1-2 (EN 60601-1-2)
1. Medical electrical equipment needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in the accompanying documents.
2. Portable and mobile RF communications equipment can affect medical electrical equipment.
3. Information regarding the cable affecting EMC is as follows.
Name
Connected Device
Maximum Length
General Specification
Network Cable
Between the Power supply
unit and the DR-ID 1200MC
20m (65.6 ft)
Cat5e or more,
UTP type and straight cable
Between the Power box and
the DR-ID 1200MC*
2m (6.56 ft)
* When the distance between the Power box and the DR-ID 1200MC is 2 meters or more, install a
commercially available HUB. In that case, use a cable with a maximum length of 2 meters between the
Power box and HUB, and a cable with a maximum length of 20 meters between HUB and
the DR-ID 1200MC.
4. The use of optional items, transducers and cables other than those specified, with the exception
of transducers and cables sold by FUJIFILM Corporation as replacement parts for internal
components, may result in increased emissions or decreased immunity of the DR-ID1270.
5. The DR-ID1270 should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the DR-ID1270 should be observed to verify normal
operation in the configuration in which it will be used.
6. Basic performance of the equipment and the system
After image data are acquired from the flat panel sensor, data correction is performed by the control
cabinet, and the image is saved in and displayed on the image processing unit.
7. Test items (Tables 1 to 4)
Table 1
Guidance and manufacturer’s declaration - electromagnetic emissions
The DR-ID1270 is intended for use in the electromagnetic environment specified below.
The customer or the user of the DR-ID1270 should assure that they are used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The DR-ID1270 uses RF energy only for their internal function.
Therefore, their RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The DR-ID1270 is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Complies
Class B
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Содержание FDR ES DR-ID1270
Страница 2: ...ii DR ID1270 Operation Manual 897N120821E...
Страница 28: ...1 22 For Safe Operation 1 DR ID1270 Operation Manual 897N120821E...
Страница 42: ...2 14 System Configuration Product Overview 2 DR ID1270 Operation Manual 897N120821E...
Страница 54: ...3 12 Basic Operation 3 DR ID1270 Operation Manual 897N120821E...
Страница 56: ...4 2 Daily Inspection and Maintenance 4 DR ID1270 Operation Manual 897N120821E...
Страница 68: ...Z 6 Appendix Z Precautions for Exposure DR ID1270 Operation Manual 897N120821E...
Страница 74: ...O 6 Appendix O Use of Optional Items DR ID1270 Operation Manual 897N120821E...
Страница 76: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...