NT 0928 Rev.A0
page : 59
The test procedure outlined below can be carried out with a 50 ml or 20 ml syringe. The operating mode
described below, reflects the flow rate Measurement software used by Fresenius Vial according to the Pr
EN 60-601-2-24 Standard for Infusion pumps. It is up to the user to adapt this procedure to the software
he uses.
5.16.1. Equipment used:
q
Scales coupled to a microcomputer
scales sensitivity (in compliance with IEC 601.1 standard, 2nd part):
flow rate value (x)
scale sensivity
x
≤
5 ml/h
1/10000e
5 ml/h < x 30
≤
ml/h
1/1000e
x > 30 ml/h
1/100e
q
Multi scales acquisition program
q
Test tube or beaker with 1 ml graduating
q
Liquid: distilled water +/ oil)
q
Luer Lock type plastic syringe (50 or 20 ml)
q
Catheter extension with Luer Lock end piece (length 100 cm, inside diameter 2.5 mm).
q
Needle:
flow rate value (x)
nodale type
x < 30 ml/h
G 26
x
≤
30 ml/h
G 18 ou G 21
5.16.2. installation
1. The equipment should be installed according to the installation drawings shown in below.
q
Important
: Make sure that the horizontal installation plane is respected.
2. Fill the syringe with 50 ml of distilled water. Prime if necessary to eliminate any air bubbles.
3. Secure the female Luer Lock end piece of the catheter extension onto the syringe and the male Luer
Lock end piece onto the needle.
4. Install the syringe onto the device.
5. Fill the test tube with water ensuring that the needle is dipped in the liquid (1 cm) and add several
drops of oil in order to create a greasy film on the surface of the liquid. In this way the user will avoid
any measurement error due to evaporation of the liquid.
6. Place the test tube in the centre of the scales platform.
7. Place the needle inside the test tube.
q
Important
: The infusion line (needle/catheter extension) must not be in contact or rest on the
Scales/test tube assembly at any time.
8. Start the device by pressing on the ON key (Injectomat 2000 Anästhesie in mains supply mode)
then prime the infusion line using the PRIME/BOLUS key.
q
Important
: check that there are no air bubbles.
Содержание INJECTOMAT 2000 Anasthesie
Страница 1: ...TECHNICAL MANUAL INJECTOMAT 2000 Anästhesie ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page 4 NT 0928 Rev A0 ...
Страница 47: ...NT 0928 Rev A0 page 43 ...
Страница 53: ...NT 0928 Rev A0 page 49 ...
Страница 54: ...Page 50 NT 0928 Rev A0 ...
Страница 55: ...NT 0928 Rev A0 page 51 ...
Страница 56: ...Page 52 NT 0928 Rev A0 ...
Страница 67: ...NT 0928 Rev A0 page 63 ...
Страница 71: ...NT 0928 Rev A0 page 67 Mechanical plunger unit Injectomat 2000 Anästhesie ...
Страница 73: ...NT 0928 Rev A0 page 69 Motor reduction bloc Mechanical kits ...
Страница 77: ...NT 0928 Rev A0 page 73 ...
Страница 78: ...Page 74 NT 0928 Rev A0 ...
Страница 79: ...NT 0928 Rev A0 8 ADDENDA ...
Страница 80: ...NT 0928 Rev A0 ...
Страница 82: ...NT 0928 Rev A0 ...