NT 0928 Rev.A0
page : 45
4.2. Wiring the components on the flexible circuit
q
Important: it is essential that the component mounting direction be respected.
1. Position the strut (1.5mm) between the 4 pin photo switch and weld the syringe head detection
photo switch flattening it against the flat cable.
Important: use a silicone between the optical switch and the flat ribbon cable. The silicon should not be
over the soldering area.
Important: verify there is no resistor continuity between the optical switch axis stop and the flat ribbon
cable.
2. Weld the disengaging switch flattening it against the flat cable.(at roughly 1.5 mm).
3. Cut the part of the flat cable corresponding to the back pressure microswitch of version A2 (see
cutting zone in the diagram below).
Weld the disengaging microswitch without flattening it against the flat cable (at roughly 1.5 mm).
4. Weld the force sensor wires
Important: when disassembling a force sensor, take care not to damage the welding pellets.
5. Pass the flat cable in the force sensor oblong.
Important: When disassembling the flexible circuit from the driving block holder, take care not
to detach or damage the holder protection square. The purposeof this square is to hold the flexible
circuit
correctly inplace and avoid short-circuits with the holder.
Disengaging switch
Diam 12 tube
Photo switch
anode long lug
Syringe head detection
Photo switch
lower marking
green wire
Содержание INJECTOMAT 2000 Anasthesie
Страница 1: ...TECHNICAL MANUAL INJECTOMAT 2000 Anästhesie ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page 4 NT 0928 Rev A0 ...
Страница 47: ...NT 0928 Rev A0 page 43 ...
Страница 53: ...NT 0928 Rev A0 page 49 ...
Страница 54: ...Page 50 NT 0928 Rev A0 ...
Страница 55: ...NT 0928 Rev A0 page 51 ...
Страница 56: ...Page 52 NT 0928 Rev A0 ...
Страница 67: ...NT 0928 Rev A0 page 63 ...
Страница 71: ...NT 0928 Rev A0 page 67 Mechanical plunger unit Injectomat 2000 Anästhesie ...
Страница 73: ...NT 0928 Rev A0 page 69 Motor reduction bloc Mechanical kits ...
Страница 77: ...NT 0928 Rev A0 page 73 ...
Страница 78: ...Page 74 NT 0928 Rev A0 ...
Страница 79: ...NT 0928 Rev A0 8 ADDENDA ...
Страница 80: ...NT 0928 Rev A0 ...
Страница 82: ...NT 0928 Rev A0 ...