Page : 30
NT 0928 Rev.A0
3.1.3.2. Syringe selection mode:
PAr.2
PAr.2
This parameter allows to choose a syringe selection mode among two:
SEL 3
= automatic selection
SEL 4
= manual selection
If
SEL 3
has been chosen and if there is more than one selectable syringe, Injectomat 2000 Anästhesie
goes automatically to "
PAr 4
PAr 4
", selectionnable syringe list configuration at the next switch on.
3.1.3.3. keyboard selectionnable infusion maximum flow rates:
PAr.3
PAr.3
.
This parameter allows to choose keyboard selectionnable infusion maximum flow rates for each type of
syringe.
1
rst
screen
Par3
Par3
50cc 150
50cc 1500
0
ml/h
ml/h
2
nd
screen
Par3
Par3
20cc 75
20cc 750
0
ml/h
ml/h
3.1.3.4. Selectionnable syringe list configuration:
PAr.4
PAr.4
This parameter allows to make out the key board selectionnable syringe list.
Example:
Braun Perfusor 50cc
Braun Perfusor 50cc
: Braun Perfusor 50cc syringe selectable.
: Braun Perfusor 50cc syringe not selectable.
MARQUE
TYPE
Braun Omnifix
20
Braun Perfusor
20
BD Perfusion
50
Braun Omnifix
50
Braun Perfusor
50
Dispomed Spritze
50
Dispomed type P
50
Fresenius Injectomat
50
Fresenius P Spritze
50
Содержание INJECTOMAT 2000 Anasthesie
Страница 1: ...TECHNICAL MANUAL INJECTOMAT 2000 Anästhesie ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page 4 NT 0928 Rev A0 ...
Страница 47: ...NT 0928 Rev A0 page 43 ...
Страница 53: ...NT 0928 Rev A0 page 49 ...
Страница 54: ...Page 50 NT 0928 Rev A0 ...
Страница 55: ...NT 0928 Rev A0 page 51 ...
Страница 56: ...Page 52 NT 0928 Rev A0 ...
Страница 67: ...NT 0928 Rev A0 page 63 ...
Страница 71: ...NT 0928 Rev A0 page 67 Mechanical plunger unit Injectomat 2000 Anästhesie ...
Страница 73: ...NT 0928 Rev A0 page 69 Motor reduction bloc Mechanical kits ...
Страница 77: ...NT 0928 Rev A0 page 73 ...
Страница 78: ...Page 74 NT 0928 Rev A0 ...
Страница 79: ...NT 0928 Rev A0 8 ADDENDA ...
Страница 80: ...NT 0928 Rev A0 ...
Страница 82: ...NT 0928 Rev A0 ...