NT 0928 Rev.A0
page : 21
2.2.1.6.1. Opto switch body syringe measurement.
q
Note: The CI opto and the obturator are specific to the Injectomat 2000 Anästhesie equipped
with "flange detector" and are not compatible with previous versions.
q
2 opto devices allow to detect 2 syringe sizes: 60 cc and 20 cc.
J2.1
opto diode +5V anode
J2.2
opto cathode common point and transistor transmitter
J2.3
opto 1 transistor collector
J2.4
opto 2 transistor collector
Control signal:
CODPT3
activate at 1
driven by microprocessor line T0.
Output signal:
SOPT4
activate at 1
register U16 D1 address $ C000
Output signal:
SOPT3
activate at 1
register U16 D0 address $ C000
.
SOPT3
SOPT4
Syringe clamp detection alarm, high position
1
0
60 cc detection
1
1
20 cc detection
0
1
Syringe clamp detection alarm, low position
0
0
The motor rotation and piston presence opto switch interfaces are located on the POWER SUPPLY
BOARD.
2.2.1.6.2. Motor rotation opto switch
Control signal:
CDOPT1
activate at 1
Output signal:
SOPT1
activate at 1
2.2.1.6.3. Piston presence opto switch (anti-siphon)
Control signal:
CDOPT2
activate at 1
Output signal:
SOPT2
activate at 1
2.2.2. Description of connectors
2.2.2.1. J1 not used
This connector is not used for the moment but it will be used for a future evolution of the software.
2.2.2.2. J2 Opto switch - syringe body connector
Pin
Description
1
ground
2
flange switch
3
opto +5 V diode anode
4
cathode diode transmitter transistor opto 1 and opto 2 common points
5
collector transistor opto 1
6
collector transistor opto 2
Содержание INJECTOMAT 2000 Anasthesie
Страница 1: ...TECHNICAL MANUAL INJECTOMAT 2000 Anästhesie ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page 4 NT 0928 Rev A0 ...
Страница 47: ...NT 0928 Rev A0 page 43 ...
Страница 53: ...NT 0928 Rev A0 page 49 ...
Страница 54: ...Page 50 NT 0928 Rev A0 ...
Страница 55: ...NT 0928 Rev A0 page 51 ...
Страница 56: ...Page 52 NT 0928 Rev A0 ...
Страница 67: ...NT 0928 Rev A0 page 63 ...
Страница 71: ...NT 0928 Rev A0 page 67 Mechanical plunger unit Injectomat 2000 Anästhesie ...
Страница 73: ...NT 0928 Rev A0 page 69 Motor reduction bloc Mechanical kits ...
Страница 77: ...NT 0928 Rev A0 page 73 ...
Страница 78: ...Page 74 NT 0928 Rev A0 ...
Страница 79: ...NT 0928 Rev A0 8 ADDENDA ...
Страница 80: ...NT 0928 Rev A0 ...
Страница 82: ...NT 0928 Rev A0 ...