Page : 44
NT 0928 Rev.A0
4 REPLACING SUB-ASSEMBLIES
q
Important: Disassemble the flexible circuit holder on the supply board before extracting the me-
chanical assembly from the lower box.
q
Important: Device operation must be completely checked after all intervention inside the device.
4.1. Mounting the flexible circuit
q
Important: Handle the flat cable with great care when mounting this sub-assembly: damage to
the flat cable will result in complete disassembly of the mechanical block.
Fitting the flat cable
1. Take the flat cable and, using the "flat cable insertion" tool, wind the flat cable on the opposite side
to the 10-point connector in the tool slot, keeping the flat cable tightened, wind it onto the equip-
ment.
2. Take the black hood of the flat cable insertion equipment and position it on the wound flat cable
part.
3. Insert the driving block cover and the input bearing on the diam. 12 tube, in the proper direction for
mounting the various components.
Important: The input bearing flange must be placed on the side external to the driving block cover.
q
Visually check that the "input b driving block cover" are properly mounted on the tube.
4. Insert the "flat cable insertion" tool in the diam. 12 tube on the input bearing side.
q
The perpendicular flat cable parts are placed in the slits found at the ends of the tube.
5. Remove the black hood and the flat cable insertion equipment.
6. Correctly position the flat cable in the two slits visible at each end of the tube.
Important: The flat cable must not be twisted inside the tube.
7. Place the flexible circuit guards at both ends of the tube, passing the flat cable between the two
holding lugs
8. Place the diam. 12 tube on the driving block making match the indexing finger of the driving block
with the hole of the tube (opposite side of the 10 point connector)
9. Match up the flexible circuit hole with the driving block holder centering tube.
10. Secure the clamping collar onto the driving block using two 2.5 x10 screws and two washers.
hood
diam.12 tube
flat cable
Содержание INJECTOMAT 2000 Anasthesie
Страница 1: ...TECHNICAL MANUAL INJECTOMAT 2000 Anästhesie ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page 4 NT 0928 Rev A0 ...
Страница 47: ...NT 0928 Rev A0 page 43 ...
Страница 53: ...NT 0928 Rev A0 page 49 ...
Страница 54: ...Page 50 NT 0928 Rev A0 ...
Страница 55: ...NT 0928 Rev A0 page 51 ...
Страница 56: ...Page 52 NT 0928 Rev A0 ...
Страница 67: ...NT 0928 Rev A0 page 63 ...
Страница 71: ...NT 0928 Rev A0 page 67 Mechanical plunger unit Injectomat 2000 Anästhesie ...
Страница 73: ...NT 0928 Rev A0 page 69 Motor reduction bloc Mechanical kits ...
Страница 77: ...NT 0928 Rev A0 page 73 ...
Страница 78: ...Page 74 NT 0928 Rev A0 ...
Страница 79: ...NT 0928 Rev A0 8 ADDENDA ...
Страница 80: ...NT 0928 Rev A0 ...
Страница 82: ...NT 0928 Rev A0 ...