2.3. DISPLAY BOARD
2.3.1. Overview
The display board is mounted directly beneath the front plate of the syringe pump. It brings together all
the facilities for operator/device dialogue: Keyboard, buzzer and display.
It is connected to the CPU by rigid connectors, forming a sandwich, with the former, held in place by
struts.
The soft keyboard is connected to the display board.
2.3.2. Functional description
The display board comprises four modules:
q
The electroluminescent display,
q
The keyboard interface,
q
The liquid crystal display,
q
The buzzer.
2.3.2.1. Electroluminescent display
The electroluminescent display is made of eighteen LED's and five 7 segment display units, with the
decimal point, except two of them which are marked (
*
) in the table below.
The diodes and display units are driven in a multiplexed, 8 segments x 8 digit matrix. The LED's and
display units are mounted with a common cathode.
The 8 segments are driven by signals SEG0 to SEG7 and the 8 digit by signals DIG0 to DIG7.
The 2 LED’s, "MAIN PRESENCE" and "FAIL" are controlled independently of the matrix.
2.3.2.1.1. LED’s table
The following table lists the various diodes used in different models.
Ref
Name
Type
Seg
Dig
LD1
Mains On
yellow
*
*
LD2
Body Alarm
red
1
0
LD3
Piston Alarm
red
2
0
LD4
ml
green
5
1
LD5
Battery
green
7
1
LD6
Fail
red
*
*
LD7
Alarm
red
4 to7
0
LD8
Worm screw 3
green
2
1
LD9
Prealarm
orange
1 to 4
2
LD10
Occlusion Alarm
red
0
0
LD11
Worm screw 2
green
3
1
LD12
Infusion End Alarm
orange
0
2
LD13
Worm screw 1
green
4
1
LD14
NU
LD15
Connection to PC
green
0
1
LD16
disengagement alarm
red
3
0
LD17
ml/h
green
5
1
LD18
Battery Alarm
red
5
2
LD19
Validation Demand
green
1
1
Содержание INJECTOMAT 2000 Anasthesie
Страница 1: ...TECHNICAL MANUAL INJECTOMAT 2000 Anästhesie ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page 4 NT 0928 Rev A0 ...
Страница 47: ...NT 0928 Rev A0 page 43 ...
Страница 53: ...NT 0928 Rev A0 page 49 ...
Страница 54: ...Page 50 NT 0928 Rev A0 ...
Страница 55: ...NT 0928 Rev A0 page 51 ...
Страница 56: ...Page 52 NT 0928 Rev A0 ...
Страница 67: ...NT 0928 Rev A0 page 63 ...
Страница 71: ...NT 0928 Rev A0 page 67 Mechanical plunger unit Injectomat 2000 Anästhesie ...
Страница 73: ...NT 0928 Rev A0 page 69 Motor reduction bloc Mechanical kits ...
Страница 77: ...NT 0928 Rev A0 page 73 ...
Страница 78: ...Page 74 NT 0928 Rev A0 ...
Страница 79: ...NT 0928 Rev A0 8 ADDENDA ...
Страница 80: ...NT 0928 Rev A0 ...
Страница 82: ...NT 0928 Rev A0 ...