2 ELECTRONIC BOARD
2.1. MOTOR POWER SUPPLY AND CONTROL BOARD
2.1.1. Functional description
In order to simplify wiring the motor power supply control board concentrates 6 functional modules,
which may be described separately:
q
power supply module,
q
motor control module,
q
analog output module,
q
disengage and anti-siphon opto switch module,
q
microswitch input module,
q
optional nurse call and RS232 interface module.
2.1.1.1. Power supply module
The power supply module consists of a cut-out power unit. It supplies all electronic components and
charges a 1,1 Ah / 1,2 Ah backup battery from a mains voltage input or a 12/15 volt DC power source. It
generates the + 5V and Vbat voltage required by the electronic components. Finally it comprises an
ON/OFF switch controlling the various power supplies.
2.1.1.1.1. Mains power supply
The mains power supply generates a DC voltage ranging from 10 and 16 volts for a maximum current of
1,2 A.
J1 mains input connector: 1 phase
2 neutral
Transformer:
TR1(see Electrical chart), 15VA output voltage: 9 Vac
Fuse protection:
F2 Principal characteristics chap1.1
Primary filtering
4.7 nf 4000 V HR capacitor, type DS1510 VDE
Secondary filtering
C10 Chemical Capacitor
Output voltage measured on TP3 for mains voltage:
230V measured (
±±
10%)
Min
Max
Unit
power off:
14
16
V dc
7 ohm charged on J4:
10
16
V dc
primary current charged:
80
mA ac
Maximum voltage
16 volts limited by the diode D 41
Minimum voltage
10 volts limited by the mains voltage - 10% and U1(MAX 652)
voltage higher than 10 volts.
Содержание INJECTOMAT 2000 Anasthesie
Страница 1: ...TECHNICAL MANUAL INJECTOMAT 2000 Anästhesie ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page 4 NT 0928 Rev A0 ...
Страница 47: ...NT 0928 Rev A0 page 43 ...
Страница 53: ...NT 0928 Rev A0 page 49 ...
Страница 54: ...Page 50 NT 0928 Rev A0 ...
Страница 55: ...NT 0928 Rev A0 page 51 ...
Страница 56: ...Page 52 NT 0928 Rev A0 ...
Страница 67: ...NT 0928 Rev A0 page 63 ...
Страница 71: ...NT 0928 Rev A0 page 67 Mechanical plunger unit Injectomat 2000 Anästhesie ...
Страница 73: ...NT 0928 Rev A0 page 69 Motor reduction bloc Mechanical kits ...
Страница 77: ...NT 0928 Rev A0 page 73 ...
Страница 78: ...Page 74 NT 0928 Rev A0 ...
Страница 79: ...NT 0928 Rev A0 8 ADDENDA ...
Страница 80: ...NT 0928 Rev A0 ...
Страница 82: ...NT 0928 Rev A0 ...