![Fresenius Vial INJECTOMAT 2000 Anasthesie Скачать руководство пользователя страница 46](http://html.mh-extra.com/html/fresenius-vial/injectomat-2000-anasthesie/injectomat-2000-anasthesie_technical-manual_2331676046.webp)
Page : 42
NT 0928 Rev.A0
3.3.15. Calibration values display
tESt.E
tESt.E
This test allows to show the calibration values stored in the EEPROM.
Every value is displayed on 3 digits.
The first line gives the shorted name of the value.
Example:
Test E LOW
Test E LOW
082
082
The keys
and
allows to pass from one channel to an other.
bat1
alarm and pre-alarm battery voltage: 6.3 V
bat2
pre-alarm battery voltage: 5.9 V
bat3
alarm battery voltage: 5.7 V
HIGH
displacement potentiometer with large 115.0 mm spacer
LOW
displacement potentiometer with small 20.0 mm spacer
0G
force meter with 0 kg
5Kg
force meter with 5 kg
3.3.16. Syringe type display
tESt.F
tESt.F
This test displays the type of syringe fitted to the Injectomat. The type is defined with the indications
given by the optos of the syringe clamp.
The type of syringe is displayed with its capacity: 20cc (20/25cc), 50cc (50/60cc). The capacity may not
be displayed if the parameters in EEPROM indicate that certain types of syringes are not included.
When the syringe clamp is in the higher and the lower position 4 dashes are displayed.
The display is constantly updated in relation to changes in the syringe clamp system.
3.3.17. Displays of the last 10 events before the last blocking error
tESt.J
tESt.J
This test allows to display the 10 last events before the last blocking error.
3.3.18. Drug library
tESt.L
tESt.L
This test allows to read the name, the author and the date of the protocol library registered in the device.
Содержание INJECTOMAT 2000 Anasthesie
Страница 1: ...TECHNICAL MANUAL INJECTOMAT 2000 Anästhesie ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page 4 NT 0928 Rev A0 ...
Страница 47: ...NT 0928 Rev A0 page 43 ...
Страница 53: ...NT 0928 Rev A0 page 49 ...
Страница 54: ...Page 50 NT 0928 Rev A0 ...
Страница 55: ...NT 0928 Rev A0 page 51 ...
Страница 56: ...Page 52 NT 0928 Rev A0 ...
Страница 67: ...NT 0928 Rev A0 page 63 ...
Страница 71: ...NT 0928 Rev A0 page 67 Mechanical plunger unit Injectomat 2000 Anästhesie ...
Страница 73: ...NT 0928 Rev A0 page 69 Motor reduction bloc Mechanical kits ...
Страница 77: ...NT 0928 Rev A0 page 73 ...
Страница 78: ...Page 74 NT 0928 Rev A0 ...
Страница 79: ...NT 0928 Rev A0 8 ADDENDA ...
Страница 80: ...NT 0928 Rev A0 ...
Страница 82: ...NT 0928 Rev A0 ...