![Fresenius Vial INJECTOMAT 2000 Anasthesie Скачать руководство пользователя страница 31](http://html.mh-extra.com/html/fresenius-vial/injectomat-2000-anasthesie/injectomat-2000-anasthesie_technical-manual_2331676031.webp)
NT 0928 Rev.A0
page : 27
3 CONFIGURATIONS, CALIBRATIONS AND CHECK
3.1. CONFIGURATIONS
3.1.1. Configuration of the pressure functions
The different operating possibilities presented will be particularly useful for adapting the syringe pump to
the specific needs of each department.
Fresenius Vial recommends the presence of its qualified personnel or of a member of the Technical De-
partment of your establishment to help you implement the configuration procedures you wish to choose.
q
N.B.: You can leave the configuration mode at any time by pressing the OFF key.
3.1.2. Moving to the pressure configuration mode
Configuration mode access is activated, when switching on, by simultaneously pressing on the keys
and
until the display
:
PrES.1
allows to scroll the parameters:
PrES.1, PrES.2, PrES.3,
etc.......... on the 7 segment screen.
allow to valid your choice and to enter in the menus.
Inside the menus, the keys:
and
, allow to display the chosen values.
q
PrES.1: Pressure limit memorization.
q
PrES.2: Maximum pressure limits.
q
PrES.3: Pressure drop detection threshold.
3.1.2.1. Pressure limit memorization
PrES.1
PrES.1
Allows to memorize the pressure limit which will be proposed every switching on.
This value is adjustable between 100 and 1100 mm Hg by step of 50 mm Hg.
Example:
Pres 1
Pres 1
950 mm Hg
950 mm Hg
If no value is memorised, the display is:
Pres 1
Pres 1
___ mm Hg
___ mm Hg
and it is the used pressure at the last switching off, which will be saved and proposed by default at the
next switching on.
Содержание INJECTOMAT 2000 Anasthesie
Страница 1: ...TECHNICAL MANUAL INJECTOMAT 2000 Anästhesie ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page 4 NT 0928 Rev A0 ...
Страница 47: ...NT 0928 Rev A0 page 43 ...
Страница 53: ...NT 0928 Rev A0 page 49 ...
Страница 54: ...Page 50 NT 0928 Rev A0 ...
Страница 55: ...NT 0928 Rev A0 page 51 ...
Страница 56: ...Page 52 NT 0928 Rev A0 ...
Страница 67: ...NT 0928 Rev A0 page 63 ...
Страница 71: ...NT 0928 Rev A0 page 67 Mechanical plunger unit Injectomat 2000 Anästhesie ...
Страница 73: ...NT 0928 Rev A0 page 69 Motor reduction bloc Mechanical kits ...
Страница 77: ...NT 0928 Rev A0 page 73 ...
Страница 78: ...Page 74 NT 0928 Rev A0 ...
Страница 79: ...NT 0928 Rev A0 8 ADDENDA ...
Страница 80: ...NT 0928 Rev A0 ...
Страница 82: ...NT 0928 Rev A0 ...