Page : 32
NT 0928 Rev.A0
3.1.3.9. Time to preventive check:
PAr.B
PAr.B
This parameter allows to choose the time to the next preventive check between 1 and 9999 continuous
running.
Par B
3500 H
When this running time is over, at the switching on, the Injectomat 2000 Anästhesie will display this
flashing message:
Par.C
It will be possible to stop this message with but it will flash again at each switching on until the check be
performed.
3.1.3.10. Memorised protocols list :
Par.C
Par.C
or
allow to scroll the parameters memorised in the EPROM.
Example:
ALFENTAN
Par.C
200.00µg/gl
If no protocol is memorised, the Injectomat 2000 Anästhesie displays:
-----------
-----------
Par.C
START
allows to start a new protocol configuration.
BOLUS
allows to erase a memorised protocol.
ALFENTANIL
Par.C
L
L
o
o
e
e
s
s
c
c
h
h
e
e
n
n
?
?
valid erasing. As long as long the erasing is not validated, pressing any key gives a
BIP
.
STOP
allows to get out
PAr.C
without memorizing the outstanding protocol configuration.
Содержание INJECTOMAT 2000 Anasthesie
Страница 1: ...TECHNICAL MANUAL INJECTOMAT 2000 Anästhesie ...
Страница 2: ......
Страница 4: ......
Страница 8: ...Page 4 NT 0928 Rev A0 ...
Страница 47: ...NT 0928 Rev A0 page 43 ...
Страница 53: ...NT 0928 Rev A0 page 49 ...
Страница 54: ...Page 50 NT 0928 Rev A0 ...
Страница 55: ...NT 0928 Rev A0 page 51 ...
Страница 56: ...Page 52 NT 0928 Rev A0 ...
Страница 67: ...NT 0928 Rev A0 page 63 ...
Страница 71: ...NT 0928 Rev A0 page 67 Mechanical plunger unit Injectomat 2000 Anästhesie ...
Страница 73: ...NT 0928 Rev A0 page 69 Motor reduction bloc Mechanical kits ...
Страница 77: ...NT 0928 Rev A0 page 73 ...
Страница 78: ...Page 74 NT 0928 Rev A0 ...
Страница 79: ...NT 0928 Rev A0 8 ADDENDA ...
Страница 80: ...NT 0928 Rev A0 ...
Страница 82: ...NT 0928 Rev A0 ...