![Fresenius Medical Care 5008 Скачать руководство пользователя страница 323](http://html.mh-extra.com/html/fresenius-medical-care/5008/5008_operating-instructions-manual_2331649323.webp)
Chapter 23: Options BLK, WET
Fresenius Medical Care 5008 OP 5/09.06
23-1
23 Options BLK, WET
23.1 To be Observed Before Using the BLK, WET Options
23.1.1 Definitions of Terms
O
BLK
Bloodline kinking and filter clotting monitoring
Detection of whether there are kinks in the blood lines between the
blood pump and the dialyzer.
Detection of whether there are clots in the dialyzer.
O
WET
Wet transductor protector
Detection of whether the venous hydrophobic filter is wet.
23.2 BLK
O
Task
If there are kinks in the blood lines or if clots are beginning to develop
in the dialyzer, this might result in a mechanically generated hemolysis.
The BLK option detects kinked blood lines between the blood pump and
the dialyzer or clots beginning to develop in the dialyzer, and emits a
warning.
O
Description
The BLK checks the extracorporeal tubing system between the blood
pump and the venous bubble catcher for changes in the flow resistance.
The BLK principle is based on monitoring the signals at the venous
pressure sensor.
The blood pump generates periodic pressure pulses. Starting at the
blood pump, these pressure pulses migrate along the tubing system
through the dialyzer and to the venous bubble catcher and are detected
in the venous pressure sensor of the dialysis system. Any change in the
pressure pulses is caused by a change in the blood flow properties in
the extracorporeal blood circuit. This, in turn, may be caused by a kink
in the tubing system or by clots developing in the dialyzer.
Содержание 5008
Страница 2: ......
Страница 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Страница 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 136: ...REINFUSION SCREEN ...
Страница 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Страница 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Страница 247: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 37 19 Prime collection bag ...
Страница 271: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 1 13 Certificates 13 1 EC Certificate ...
Страница 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Страница 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 278: ...OCM SCREEN OCM diagram ...
Страница 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 286: ...ONLINE SCREEN ...
Страница 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Страница 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 292: ...SINGLE NEEDLE SCREEN ...
Страница 293: ...Chapter 17 Option SN Single Needle Fresenius Medical Care 5008 OP 5 09 06 17 1 17 Option SN Single Needle ...
Страница 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 305: ...Chapter 19 Option CBPM Fresenius Medical Care 5008 OP 5 09 06 19 1 19 Option CBPM not yet available ...
Страница 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 308: ...Inserting the tubing system into the measuring heads BTM events a b BTM SCREEN BTM T control graphic BTM rec graphic ...
Страница 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...