Chapter 11: System Description
Fresenius Medical Care 5008 OP 5/09.06
11-49
4.3
OCM calibration performed.
–
T
5
Extracorporeal components
5.1
Arterial pressure display
5.1.1
Zero point of arterial pressure display checked (standby operation)
Desired value: 0 mmHg; –5 mmHg to +5 mmHg
–
________
T
5.1.2
Slope of arterial pressure display tested. (standby operation)
Desired value: 300 mmHg; –5 mmHg to +5 mmHg
–
________
T
5.2
Venous pressure display
5.2.1
Zero point of venous pressure display checked (standby operation)
Desired value: 0 mmHg; –5 mmHg to +5 mmHg
–
________
T
5.2.2
Slope of venous pressure display tested. (standby operation)
Desired value: 300 mmHg; –5 mmHg to +5 mmHg
–
________
T
5.3
Venous clamp checked.
A change in pressure must not exceed the following values within
3 minutes:
Arterial pressure display, maximum change in pressure:
±
5 mmHg
Pressure display of reference measuring instrument, maximum
pressure drop: –0.1 bar
–
T
6
Options
6.1
BPM
6.1.4
TMC
Leakage test performed.
Pressure leakage rate: <6 mmHg/min.
–
________
T
6.1.5
TMC
Safety valve tested.
System emptied at 320 mmHg,
±
10 mmHg
–
________
T
6.1.6
TMC
Blood pressure measurement performed.
Measured values are plausible.
–
T
6.1.7
TMC
Calibration performed.
Pressure values / tolerance
250 mmHg / ±3 mmHg
200 mmHg / ±3 mmHg
150 mmHg / ±3 mmHg
100 mmHg / ±3 mmHg
50 mmHg / ±3 mmHg
–
System
/
ref.
____/____
____/____
____/____
____/____
____/____
T
7
Electrical safety
In Germany according to DIN VDE 0751-1, edition 10/2001.
In other countries, observe the local regulations!
7.1
Visual inspections performed according to item 1.
T
7.2
Protective earth resistance measured.
Max. 0.3
Ω
(with power cable)
______
Ω
T
No.
Description
Correct.
Meas.
value
Содержание 5008
Страница 2: ......
Страница 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Страница 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 136: ...REINFUSION SCREEN ...
Страница 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Страница 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Страница 247: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 37 19 Prime collection bag ...
Страница 271: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 1 13 Certificates 13 1 EC Certificate ...
Страница 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Страница 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 278: ...OCM SCREEN OCM diagram ...
Страница 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 286: ...ONLINE SCREEN ...
Страница 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Страница 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 292: ...SINGLE NEEDLE SCREEN ...
Страница 293: ...Chapter 17 Option SN Single Needle Fresenius Medical Care 5008 OP 5 09 06 17 1 17 Option SN Single Needle ...
Страница 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 305: ...Chapter 19 Option CBPM Fresenius Medical Care 5008 OP 5 09 06 19 1 19 Option CBPM not yet available ...
Страница 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 308: ...Inserting the tubing system into the measuring heads BTM events a b BTM SCREEN BTM T control graphic BTM rec graphic ...
Страница 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...