Chapter 11: System Description
Fresenius Medical Care 5008 OP 5/09.06
11-3
11.1.6 Guidance and Manufacturer's Declaration on EMC (IEC 60601-1-2:2001)
O
Electromagnetic emissions
O
Electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic emissions
The 5008 hemodialysis system is intended for use in the electromagnetic environment specified below. The
customer or the user of the 5008 hemodialysis system should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The 5008 hemodialysis system uses RF energy only for its
internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
The 5008 hemodialysis system is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The 5008 hemodialysis system is intended for use in the electromagnetic environment specified below. The
customer or the user of the 5008 hemodialysis system should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
±2 kV for power
supply lines
Not applicable
Mains power quality should be that of a
typical commercial and/or hospital
environment.
Surge IEC 61000-4-5 ±1 kV differential
mode
±2 kV common mode
±1 kV differential
mode
±2 kV common mode
Mains power quality should be that of a
typical commercial and/or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5 % U
T
(>95 % dip
in
U
T
) for 0.5 cycle
<5 % U
T
(>95 % dip
in
U
T
) for 0.5 cycle
After power supply interruptions, the 5008
hemodialysis system battery takes over
the supply without delay.
40 % U
T
(60 % dip in
U
T
) for 5 cycles
40 % U
T
(60 % dip in
U
T
) for 5 cycles
70 % U
T
(30 % dip in
U
T
) for 25 cycles
70 % U
T
(30 % dip in
U
T
) for 25 cycles
<5 % U
T
(>95 % dip
in
U
T
) for 5 sec
<5 % U
T
(>95 % dip
in
U
T
) for 5 sec
Содержание 5008
Страница 2: ......
Страница 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Страница 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 136: ...REINFUSION SCREEN ...
Страница 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Страница 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Страница 247: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 37 19 Prime collection bag ...
Страница 271: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 1 13 Certificates 13 1 EC Certificate ...
Страница 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Страница 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 278: ...OCM SCREEN OCM diagram ...
Страница 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 286: ...ONLINE SCREEN ...
Страница 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Страница 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 292: ...SINGLE NEEDLE SCREEN ...
Страница 293: ...Chapter 17 Option SN Single Needle Fresenius Medical Care 5008 OP 5 09 06 17 1 17 Option SN Single Needle ...
Страница 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 305: ...Chapter 19 Option CBPM Fresenius Medical Care 5008 OP 5 09 06 19 1 19 Option CBPM not yet available ...
Страница 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 308: ...Inserting the tubing system into the measuring heads BTM events a b BTM SCREEN BTM T control graphic BTM rec graphic ...
Страница 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...