Chapter 10: Other Functions
Fresenius Medical Care 5008 OP 5/09.06
10-31
The circulation time is 10 minutes. During circulation time, the
temperature of the blood in the blood line changes. The Circulation
function can be repeated as often as desired and can be aborted at any
time.
In the
OPTIONS
MENU (or directly, if defined as an options button)
Touch the
CIRCULATION
button.
Touch the Circulation
Start
button.
Message:
Extracorporeal circuit will be recirculated! –
Circulation
–
Dialysis
Continue
Use an adapter to interconnect the arterial and the venous patient
access lines.
Touch the
Circulation
button.
Blood flow 100 ml/min
Message:
Stop circulation? –
OK
If the patient has to be reconnected, touch the
OK
button.
Message:
Has the patient been reconnected?
Circulation
– Dialysis
Continue
Reconnect the patient.
Touch the
Continue
button.
Check the blood flow.
10.11 Setting the Level in the Venous Bubble Catcher
The venous bubble catcher must be correctly inserted into the level
detector. Mind the locator for the venous bubble catcher.
In the
BLOOD SYSTEM
menu
Setting the level with the
V
or
W
button
Press and hold the
V
or
W
button until the level has reached the
desired position.
Setting the level with the
Level Set button
Touch the
Level Set
button.
(The function is available only if the rinse procedure using the rinse
volume has been completed.)
The level is automatically lowered and then raised to the correct
position.
Содержание 5008
Страница 2: ......
Страница 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Страница 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 136: ...REINFUSION SCREEN ...
Страница 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Страница 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Страница 247: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 37 19 Prime collection bag ...
Страница 271: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 1 13 Certificates 13 1 EC Certificate ...
Страница 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Страница 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 278: ...OCM SCREEN OCM diagram ...
Страница 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 286: ...ONLINE SCREEN ...
Страница 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Страница 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 292: ...SINGLE NEEDLE SCREEN ...
Страница 293: ...Chapter 17 Option SN Single Needle Fresenius Medical Care 5008 OP 5 09 06 17 1 17 Option SN Single Needle ...
Страница 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 305: ...Chapter 19 Option CBPM Fresenius Medical Care 5008 OP 5 09 06 19 1 19 Option CBPM not yet available ...
Страница 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 308: ...Inserting the tubing system into the measuring heads BTM events a b BTM SCREEN BTM T control graphic BTM rec graphic ...
Страница 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...