Chapter 6: Treatment
Fresenius Medical Care 5008 OP 5/09.06
6-1
6
Treatment
6.1
TreatmentusingONLINE
plus
™
Irrespective of the treatment mode, all 5008 ONLINE
plus
™
hemodialysis systems can be operated without rinse or infusion
solutions provided in NaCl bags. The fluid volumes required for
preparation, bolus administration or during reinfusion will then be
produced ONLINE by the 5008 hemodialysis system according to the
actual requirements, thus saving both cost and time.
6.1.1
Connecting the Patient — Double-Needle Dialysis Using ONLINE
plus
™
PREPARATION SCREEN
Touch the
Blood pump I/O
button. (
Blood pump I/O
indicator grey.)
The blood pump stops.
Message:
Priming / Rinsing –
Continue
–
Exit
Touch the
Exit
button.
Disconnect the arterial tubing system from the SafeLine™.
Connect the SafeLine™ before (predilution) or below the dialyzer
(postdilution).
Connect the arterial line to the patient.
Remove the rinse connector.
Disconnect the venous tubing system from the rinse connector and
connect it to the patient or, if desired, to a separate prime collection bag.
Close the rinse port.
Message:
Start the blood pump –
Confirm
Touch the
Confirm
button.
Optical detector sensed blood.
Message:
Blood detected - Dialysis
Start
The blood pump stops. The arterial/venous occlusion clamp closes.
Mute
LED is flashing.
Connect the venous connector of the tubing system to the patient
(provided this has not been done yet).
Caution
The patient may only be connected if the following conditions are
fulfilled:
The header bar of the screen has to be grey (normal operating mode).
The hemodialysis system must be alarm-free.
Air-free priming of the extracorporeal circuit has been completed.
Содержание 5008
Страница 2: ......
Страница 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Страница 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 136: ...REINFUSION SCREEN ...
Страница 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Страница 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Страница 247: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 37 19 Prime collection bag ...
Страница 271: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 1 13 Certificates 13 1 EC Certificate ...
Страница 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Страница 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 278: ...OCM SCREEN OCM diagram ...
Страница 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 286: ...ONLINE SCREEN ...
Страница 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Страница 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 292: ...SINGLE NEEDLE SCREEN ...
Страница 293: ...Chapter 17 Option SN Single Needle Fresenius Medical Care 5008 OP 5 09 06 17 1 17 Option SN Single Needle ...
Страница 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 305: ...Chapter 19 Option CBPM Fresenius Medical Care 5008 OP 5 09 06 19 1 19 Option CBPM not yet available ...
Страница 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 308: ...Inserting the tubing system into the measuring heads BTM events a b BTM SCREEN BTM T control graphic BTM rec graphic ...
Страница 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Страница 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...