93
14.3
Volume To Be Infused (VTBI)
Applicable for all syringe sizes (50/60 mL, 30 mL, 20 mL, 10 mL, 5 mL)
14.4
Dose To Be Infused (DTBI)
Applicable for all syringe sizes (50/60 mL, 30 mL, 20 mL, 10 mL, 5 mL)
14.5
Infusion Time
Applicable for all syringe sizes (50/60 mL, 30 mL, 20 mL, 10 mL, 5 mL)
Format
Range of Settings
Default
Value
Minimum
Increment
Volume Limit
mL
0.1
999
0.1
0.1
1
(0.1
99.9)
(100
999)
Volume/Time
mL
0.1
99.9
0.1
0.1
Direct Bolus
mL
0.1
50
N/A
0.1
Programmed Bolus
mL
0.1
99.9
0.1
0.1
Format
Range of Settings
Default
Value
Minimum
Increment
Dose Rate
Unit
0.01
9999
N/A
0.001
0.01
0.1
1
(0.010
4.999)
(5.00
9.99)
(10.0
99.9)
(100
9999)
Programmed Bolus
Unit
0.01
9999
N/A
0.01
0.1
1
(0.01
9.99)
(10
99.9)
(100
9999)
Loading Dose
Unit
0.01
9999
N/A
0.01
0.1
1
(0.01
9.99)
(10.0
99.9)
(100
9999)
Format
Range of Settings
Default
Value
Minimum
Increment
Infusion Rate
_h_min_s
00h00min01s
96h00min00s
N/A
00h00min01s
Programmed Bolus
_h_min_s
00h00min01s
24h00min00s 00h02min00s
00h00min01s
Loading Dose
_h_min_s
00h00min01s
24h00min00s 00h02min00s
00h00min01s
KVO Silence Alarm
Duration
_h_min
00h01min
12h00min
00h01min
00h01min
Pause*
_h_min
00h01min
24h00min
00h01min
00h01min