11.4.10 Audio-only Information Signals
In the case of a system malfunction, the alarm sounds and an error message
Erxx(yyyy) !!!
is displayed.
1.
Record the error message
Erxx(yyyy) !!!
.
2.
Disconnect the pump from the power supply.
3.
Reset the device by pressing the
key for 10-15 seconds, and release it
when the device powers off.
Warning
If the alarm persists when the pump is powered on again, do not use the
device on a patient, and contact qualified biomedical engineering staff in
your healthcare facility or your
Fresenius Kabi
sales representative.
Type
Comment
Stops
Infusion?
Activation
Switch mode
Beep until key is
released
No
Beep starts when action is not allowed
End of loading dose
3 beeps
No
At the end of the loading dose
End of programmed
bolus
3 beeps
No
At the end of a programmed bolus
Start infusion at the end
of pause
3 beeps
N/A
At the end of a pause, when the infusion
automatically starts
AC power connection
1 beep
No
When power is connected
Forbidden key
1 beep
No
Repeated until key is released
Key beep
1 beep
No
For each key pressed
Other non validation
beep
1 beep
No
For each key pressed
Manual bolus
1 beep
No
Repeated for each mL infused
Syringe prime
1 beep
N/A
Repeated every 5 seconds
End of pause
4 beeps
N/A
At the end of pause - repeated